首页> 外文期刊>Medical Physics >Dose distribution close to metal implants in Gamma Knife Radiosurgery: a Monte Carlo study.
【24h】

Dose distribution close to metal implants in Gamma Knife Radiosurgery: a Monte Carlo study.

机译:伽马刀放射外科中金属植入物附近的剂量分布:蒙特卡洛研究。

获取原文
获取原文并翻译 | 示例
           

摘要

Materials with high atomic numbers favor the occurrence of the photoelectric effect when they are irradiated with gamma rays. Therefore, the photoelectric effects of metal implants within the target regions in Gamma Knife Radiosurgery are worth studying. In the present work, Monte Carlo simulations using EGS4 were employed to investigate the resulting dose enhancements. A dose enhancement as high as 10% was observed close to a platinum implant along the x and y axes, while no significant dose enhancements were observed for silver, stainless steel 301, and titanium ones. A dose enhancement as high as 20% was observed close to the platinum implant along the z axis at the superior position of the metal-phantom interface and was 10% higher for other metal implants.
机译:具有高原子序数的材料在被伽马射线辐照时有利于光电效应的发生。因此,伽马刀放射外科目标区域内金属植入物的光电效应值得研究。在当前的工作中,采用了使用EGS4的蒙特卡洛模拟来研究剂量的增加。在沿x和y轴靠近铂植入物的位置,观察到剂量增加高达10%,而对于银,不锈钢301和钛,没有观察到明显的剂量增加。在金属-幻像界面的优越位置,沿着z轴靠近铂植入物,观察到剂量增加高达20%,而对于其他金属植入物,则增加了10%。