首页> 外文期刊>Medical Physics >The thermal characteristics of different diodes on in vivo patient dosimetry.
【24h】

The thermal characteristics of different diodes on in vivo patient dosimetry.

机译:体内患者剂量测定中不同二极管的热特性。

获取原文
获取原文并翻译 | 示例
           

摘要

Diode sensitivity variations with temperature (SVWT) have been reported to vary from small negative values up to 0.6% per degrees C. Thus it is possible for diode calibration factors established at room temperature (approximately 20 degrees C) to yield errors in the range of -1% to +9% when diodes are placed on a patient's skin (approximately 30 degrees C) for in vivo entrance dose measurements. In this study we simulated several skin temperatures using a temperature-controlled aluminum surface in contact with a section of Bolus. The internal temperatures of several diodes with different buildup thickness were monitored as a function of time when placed in contact with the heated bolus. Our results indicate that for different combinations of room temperature (18 degrees C-23 degrees C) and patient skin temperature (28 degrees C-34 degrees C) diodes reached 90% of their equilibrium temperature within 3-5 min. In addition, the range of typical skin temperatures was determined by measurements performed on a number of actual patients under clinical conditions. Based on the results of our experiments a protocol was developed to minimize the temperature based errors for in vivo dosimetry.
机译:据报道,二极管灵敏度随温度的变化(SVWT)从很小的负值变化到每摄氏度0.6%。因此,在室温(约20摄氏度)下建立的二极管校准因子可能会产生范围为当将二极管放置在患者皮肤(大约30摄氏度)上进行体内入口剂量测量时,为-1%至+ 9%。在这项研究中,我们使用与部分Bolus接触的温度受控的铝表面模拟了几种皮肤温度。当放置在与加热的大剂量药丸接触时,随时间变化监测几个具有不同堆积厚度的二极管的内部温度。我们的结果表明,对于室温(18摄氏度至23摄氏度)和患者皮肤温度(28摄氏度至34摄氏度)的不同组合,二极管在3-5分钟内达到其平衡温度的90%。另外,典型的皮肤温度范围是通过在临床条件下对许多实际患者进行测量而确定的。基于我们的实验结果,开发了一种方案以最小化体内剂量法中基于温度的误差。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号