Monte Carlo calculations and experimental measurements of dosimetry parameters of a new 103Pd source.

机译:新型103Pd源的剂量学参数的蒙特卡罗计算和实验测量。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Permanent prostate implantation using 125I (iodine) or 103Pd (palladium) sources is a popular treatment option in the management of early prostate cancer. As sources of new designs are developed and marketed for application in permanent prostate implantations, their dosimetric characteristics must be carefully determined in order to maintain the accuracy of patient treatment. This report presents the results of experimental measurements and Monte Carlo calculations of the dosimetric parameters performed for a newly available 103Pd seed source. The measurements were performed in a large scanning water phantom using a diode detector. The positioning of the source and detector was achieved by a computer-controlled positioning mechanism in the scanning water phantom. The dose rate constant in water for the new 103Pd source was determined from measurements with the diode detector calibrated with 125I sources of known air-kerma strength. The radial dose function values for the source were measured using the diode detector. Monte Carlo photon transport calculations were then used to calculate the dosimetric parameters of dose rate constant, radial dose function, and anisotropy function using an accurate geometric model of the source. The measured dose rate constant of 0.693 cGy/U-hr compares well with the Monte Carlo calculated value of 0.677 cGy/U-hr. These results are further compared with data on existing 103Pd sources.
机译:在早期前列腺癌的治疗中,使用125 I(碘)或103 Pd(钯)源进行永久性前列腺植入是一种流行的治疗选择。随着新设计来源的开发和市场化以用于永久性前列腺植入,必须仔细确定其剂量特性,以保持患者治疗的准确性。该报告介绍了针对新近可用的103Pd种子源进行的剂量学参数的实验测量和蒙特卡洛计算的结果。使用二极管检测器在大型扫描水体模中进行测量。通过计算机控制的定位机构在扫描水体模中实现源和探测器的定位。对于新的103Pd离子源,水中的剂量率常数是根据二极管探测器的测量结果确定的,该二极管探测器用已知的空气动能强度的125I离子源进行了校准。使用二极管检测器测量放射源的径向剂量函数值。然后,使用精确的源几何模型,使用蒙特卡洛光子传输计算来计算剂量率常数,径向剂量函数和各向异性函数的剂量学参数。测得的剂量率常数为0.693 cGy / U-hr,与Monte Carlo计算值0.677 cGy / U-hr相当。将这些结果与现有103Pd来源的数据进行了进一步比较。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号