Dependence of linac output on the switch rate of an intensity-modulated tomotherapy collimator.

机译:直线加速器输出对强度调制断层扫描准直仪的转换速率的依赖性。

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

The electro-mechanical, multivane intensity modulated collimator ("MIMiC") slit collimator with 40 vanes has been applied in the delivery of inversely planned sequential tomotherapy to over 4,000 patients. The collimator is binary in that each vane switches between fully open or closed status. Resulting beamlet patterns provide the intensity distributions imparting dose to the patient. The bouncing and damping of vanes at the two ends of their travel cause transient dose perturbations near and at the borders of the treatment field. These perturbations are not explicitly modeled by the planning system. Clinical beamlet profiles and output factors may then differ from those in the planning system and as a function of the vane switch period. A mechanical model of vane switching was developed to describe this dependency. Dose output and distribution of seven simple vane patterns with different switch times were measured with ionization chambers and radiographic films in polystyrene and anthropomorphic phantoms. Linac output dependence on switch time relative to vane open time was determined for four intensity modulated radiotherapy (IMRT) patients from measurements of an ionization chamber embedded in a cylindrical polystyrene phantom. Results demonstrate output dependence on switch time and, accordingly, on the servo mechanism for monitor units, arc length, dose rate, and gantry speed. In conclusion, the output dependence borders on clinical significance-improvements to collimator, dose calculation, commissioning, and quality assurance (QA) are suggested.
机译:带有40个叶片的机电式多叶片强度调制准直仪(“ MIMiC”)狭缝准直仪已用于向4,000多名患者提供反向计划的连续tomotherapy。准直器是二进制的,因为每个叶片在完全打开或关闭状态之间切换。所得的子束图案提供将剂量传递给患者的强度分布。叶片行进的两端的弹跳和阻尼会在治疗区域的边界附近和边界处引起短暂的剂量扰动。规划系统未明确建模这些扰动。然后,临床子束轮廓和输出因子可能会与计划系统中的不同,并且取决于叶片切换周期。开发了叶片切换的机械模型来描述这种依赖性。用电离室和射线照相胶片在聚苯乙烯和拟人模型中测量了七个具有不同切换时间的简单叶片模式的剂量输出和分布。通过对嵌入圆柱状聚苯乙烯体模中的电离室的测量,确定了四位调强放射治疗(IMRT)患者的直线加速器输出相对于切换时间(相对于叶片开启时间)的依赖性。结果表明输出取决于开关时间,并因此取决于监控器的伺服机构,电弧长度,剂量率和机架速度。总之,建议对输出的依赖性与准直仪的临床意义改善,剂量计算,调试和质量保证(QA)密切相关。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号