A study of experimental measurements of dosimetric parameters in HDR IR-192 source.

机译:HDR IR-192源中剂量参数的实验测量研究。

获取原文
获取原文并翻译 | 示例
           

摘要

Thermoluminescent dosimeters have been used to perform dosimetry measurements for the widely used Ir-192 microSelectron-HDR source with an improved polystyrene phantom. Radial dose functions and anisotropy functions, main parameters of 2D dose-rate formalism from the TG-43U1 protocol, have been obtained experimentally. Measurement results are compared with that of the Monte Carlo calculations reported, and no difference has been found between them.
机译:热发光剂量计已用于对带有改进的聚苯乙烯体模的广泛使用的Ir-192 microSelectron-HDR光源进行剂量测定。径向剂量函数和各向异性函数是TG-43U1协议中二维剂量率形式的主要参数,已通过实验获得。将测量结果与所报告的蒙特卡洛计算结果进行比较,发现两者之间没有差异。