The PTW-Linacheck was assessed for its ability to monitor linear accelerator radiation output constancy. The key issues that were considered were the setup for daily output measurements, e.g., requirements for build-up and backscatter material, and the reproducibility and linearity of the device with linear accelerator output. An appropriate measurement setup includes a 10 x 10 cm field at 100 cm FSD, 5 cm backscatter, and no added build-up for 4 MeV electron beams, 1 cm added build-up for 6-16 MeV electron beams and 5 cm added build-up for 6-15 MV photon beams. Using this measurement setup, the dose linearity and short-term reproducibility were acceptable; however, the Linacheck should be recalibrated on a monthly basis to ensure acceptable long-term reproducibility.
机译:评估了PTW-Linacheck监测线性加速器辐射输出恒定性的能力。所考虑的关键问题是日常输出测量的设置,例如对堆积和反向散射材料的要求,以及具有线性加速器输出的设备的可再现性和线性。适当的测量设置包括:在100 cm FSD时有10 x 10 cm的视野,5 cm的反向散射,对于4 MeV电子束没有增加的累积,对于6-16 MeV电子束没有增加的累积和5 cm累积的增加-用于6-15 MV光子束。使用这种测量设置,剂量线性和短期可重复性是可以接受的。但是,应每月重新校准Linacheck,以确保可接受的长期再现性。