首页> 外文期刊>Medical and Biological Engineering and Computing: Journal of the International Federation for Medical and Biological Engineering >Investigation of photoplethysmographic signals and blood oxygen saturation values on healthy volunteers during cuff-induced hypoperfusion using a multimode PPG/SpO 2 sensor
【24h】

Investigation of photoplethysmographic signals and blood oxygen saturation values on healthy volunteers during cuff-induced hypoperfusion using a multimode PPG/SpO 2 sensor

机译:使用多模式PPG / SpO 2传感器对健康志愿者袖带性灌注不足期间的光电容积描记信号和血氧饱和度值进行调查

获取原文
获取原文并翻译 | 示例
           

摘要

Photoplethysmography (PPG) is a technique widely used to monitor volumetric blood changes induced by cardiac pulsations. Pulse oximetry uses the technique of PPG to estimate arterial oxygen saturation values ( (SpO 2)). In poorly perfused tissues, (SpO 2) readings may be compromised due to the poor quality of the PPG signals. A multimode finger PPG probe that operates simultaneously in reflectance, transmittance and a combined mode called "transreflectance" was developed, in an effort to improve the quality of the PPG signals in states of hypoperfusion. Experiments on 20 volunteers were conducted to evaluate the performance of the multimode PPG sensor and compare the results with a commercial transmittance pulse oximeter. A brachial blood pressure cuff was used to induce artificial hypoperfusion. Results showed that the amplitude of the transreflectance AC PPG signals were significantly different (p 0.05) than the AC PPG signals obtained from the other two conventional PPG sensors (reflectance and transmittance). At induced brachial pressures between 90 and 135 mmHg, the reflectance finger pulse oximeter failed 25 times (failure rate 42.2 %) to estimate SpO 2 values, whereas the transmittance pulse oximeter failed 8 times (failure rate 15.5 %). The transreflectance pulse oximeter failed only 3 times (failure rate 6.8 %) and the commercial pulse oximeter failed 17 times (failure rate 29.4 %).
机译:光电容积描记术(PPG)是一种广泛用于监测由心脏搏动引起的体积血液变化的技术。脉搏血氧饱和度测定法使用PPG技术估算动脉血氧饱和度值((SpO 2))。在灌注不良的组织中,由于Spg信号质量较差,(SpO 2)读数可能会受到影响。为了提高低灌注状态下PPG信号的质量,开发了一种同时在反射率,透射率和称为“透反射”的组合模式下工作的多模手指PPG探头。对20名志愿者进行了实验,以评估多模式PPG传感器的性能,并将结果与​​商业透射脉冲血氧仪进行比较。臂式血压袖带用于诱发人工灌注不足。结果表明,透反射AC PPG信号的幅度与从其他两个常规PPG传感器获得的AC PPG信号(反射率和透射率)显着不同(p <0.05)。在90到135 mmHg的诱导肱动脉压下,反射手指脉搏血氧仪无法估计25次SpO 2值(失败率42.2%),而透射脉搏血氧仪8次失败(失败率15.5%)。透反射式脉搏血氧仪仅发生3次故障(故障率6.8%),商用脉搏血氧仪发生17次故障(故障率29.4%)。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号