An optimized online verification imaging procedure for external beam partial breast irradiation.

机译:优化的在线验证成像程序,用于外部束部分乳房照射。

获取原文
获取原文并翻译 | 示例
           

摘要

The purpose of this study was to evaluate the capabilities of a kilovoltage (kV) on-board imager (OBI)-equipped linear accelerator in the setting of on-line verification imaging for external-beam partial breast irradiation. Available imaging techniques were optimized and assessed for image quality using a modified anthropomorphic phantom. Imaging dose was also assessed. Imaging techniques were assessed for physical clearance between patient and treatment machine using a volunteer. Nonorthogonal kV image pairs were identified as optimal in terms of image quality, clearance, and dose. After institutional review board approval, this approach was used for 17 patients receiving accelerated partial breast irradiation. Imaging was performed before every fraction verification with online correction of setup deviations >5 mm (total image sessions = 170). Treatment staff rated risk of collision and visibility of tumor bed surgical clips where present. Image session duration and detected setup deviations were recorded. For all cases, both image projections (n = 34) had low collision risk. Surgical clips were rated as well as visualized in all cases where they were present (n = 5). The average imaging session time was 6 min, 16 sec, and a reduction in duration was observed as staff became familiar with the technique. Setup deviations of up to 1.3 cm were detected before treatment and subsequently confirmed offline. Nonorthogonal kV image pairs allowed effective and efficient online verification for partial breast irradiation. It has yet to be tested in a multicenter study to determine whether it is dependent on skilled treatment staff.
机译:这项研究的目的是评估在体外部分乳房照射的在线验证成像设置中,配备一千伏(kV)车载成像器(OBI)的线性加速器的功能。优化的可用成像技术,并使用改进的拟人化体模对图像质量进行评估。还评估了成像剂量。使用志愿者评估了成像技术在患者和治疗机之间的物理间隙。在图像质量,清除率和剂量方面,非正交kV图像对被确定为最佳。经机构审查委员会批准后,该方法用于17例接受部分乳房加速照射的患者。在每次馏分验证之前都要进行成像,并在线校正设置偏差> 5 mm(总成像时间= 170)。治疗人员对发生碰撞的风险和存在的肿瘤床手术夹的可见性进行了评估。记录图像会话持续时间和检测到的设置偏差。对于所有情况,两个图像投影(n = 34)的碰撞风险都很低。在所有存在手术夹的情况下,对手术夹进行评级并可视化(n = 5)。平均成像时间为6分16秒,随着工作人员熟悉该技术,观察到的持续时间减少了。在治疗之前检测到最大1.3厘米的设置偏差,然后离线确认。非正交kV图像对允许对部分乳房照射进行有效且有效的在线验证。它尚未在多中心研究中进行测试,以确定它是否依赖于熟练的治疗人员。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号