We used a second treatment planning system (TPS) for independent verification of the dose calculated by our primary TPS in the context of patient-specific quality assurance (QA) for intensity-modulated radiation therapy (IMRT). QA plans for 24 patients treated with inverse planned dynamic IMRT were generated using the Nomos Corvus TPS. The plans were calculated on a computed tomography scan of our QA phantom that consists of three Solid Water slabs sandwiching radiochromic films, and an ion chamber that is inserted into the center slab of the phantom. For the independent verification, the dose was recalculated using the Varian Eclipse TPS using the multileaf collimator files and beam geometry from the original plan. The data was then compared in terms of absolute dose to the ion chamber volume as well as relative dose on isodoses calculated at the film plane. The calculation results were also compared with measurements performed for each case. When comparing ion chamber doses, the mean ratio was 0.999 (SD 0.010) for Eclipse vs. Corvus, 0.988 (SD 0.020) for the ionization chamber measurements vs. Corvus, and 0.989 (SD 0.017) for the ionization chamber measurements vs. Eclipse. For 2D doses with gamma histogram, the mean value of the percentage of pixels passing the criteria of 3%, 3 mm was 94.4 (SD 5.3) for Eclipse vs. Corvus, 85.1 (SD 10.6) for Corvus vs. film, and 93.7 (SD 4.1) for Eclipse vs. film; and for the criteria of 5%, 3 mm, 98.7 (SD 1.5) for Eclipse vs. Corvus, 93.0 (SD 7.8) for Corvus vs. film, and 98.0 (SD 1.9) for Eclipse vs. film. We feel that the use of the Eclipse TPS as an independent, accurate, robust, and time-efficient method for patient-specific IMRT QA is feasible in clinic.
机译:我们使用第二个治疗计划系统(TPS)对由主要TPS计算出的剂量进行独立验证,以进行强度调制放射治疗(IMRT)的患者特定质量保证(QA)。使用Nomos Corvus TPS制定了24例接受反向计划动态IMRT治疗的患者的QA计划。该计划是根据对我们的QA体模的计算机断层扫描技术计算得出的,该体模体由三个夹有放射致变色膜的固体水平板和一个插入到体模中央平板中的离子室组成。为了进行独立验证,使用多叶准直仪文件和原始计划的射束几何形状,使用Varian Eclipse TPS重新计算了剂量。然后将数据与离子室体积的绝对剂量以及在膜平面上计算的同工位上的相对剂量进行比较。还将计算结果与每种情况下执行的测量进行比较。在比较离子室剂量时,Eclipse与Corvus的平均比率为0.999(SD 0.010),电离室测量与Corvus的平均值为0.988(SD 0.020),电离室测量与Eclipse的平均值为0.989(SD 0.017)。对于具有γ直方图的2D剂量,通过Eclipse和Corvus的3%,3毫米标准的像素百分比的平均值分别为94.4(SD 5.3),Corvus与胶片的85.1(SD 10.6)和93.7( SD 4.1),用于Eclipse与电影;对于Eclipse对Corvus而言,标准值为5%,3毫米,98.7(SD 1.5),对于Corvus对胶片而言为93.0(SD 7.8),对于Eclipse对胶卷而言为98.0(SD 1.9)。我们认为将Eclipse TPS用作针对患者特定的IMRT QA的独立,准确,可靠且省时的方法在临床上是可行的。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号