首页> 外文期刊>Medical dosimetry: official journal of the American Association of Medical Dosimetrists >Evaluation of a 3D diode array dosimeter for helical tomotherapy delivery QA.
【24h】

Evaluation of a 3D diode array dosimeter for helical tomotherapy delivery QA.

机译:评估3D二极管阵列剂量计用于螺旋断层扫描质量保证。

获取原文
获取原文并翻译 | 示例
           

摘要

The Delta4 biplanar diode array dosimeter was validated for helical tomotherapy delivery QA. The basic detector characteristics were found to be satisfactory in terms of short-term reproducibility (0.1%), linearity (<0.1%), dose rate dependence (0.4%), and absolute calibration accuracy (0.4% in the center of the phantom compared with the independently calibrated diode). Relative calibration of the arrays was verified by comparison with film and by rotating the detector 180 degrees . The dosimeter response to rotational irradiation changed by no more than 0.2% when one of the detector boards was replaced by the homogeneous phantom material. The daily output correction factor can be derived from a Delta4 measurement in a uniform cylindrical field. The gamma(3%, 3 mm) passing rate (absolute dose) was above 90% for all 9 evaluated clinical plans, and above 96% for all but one. The mean passing rate was 97 +/- 2.7%. The plans varied in modulation factor, pitch, and calculation grid size. For best results, the phantom needs to be aligned carefully, preferably by megavoltage computed tomography imaging.
机译:Delta4双平面二极管阵列剂量计经验证可用于螺旋层析疗法QA。发现基本的检测器特性在短期再现性(0.1%),线性(<0.1%),剂量率依赖性(0.4%)和绝对校准准确度(在幻像中心为0.4%)方面令人满意与独立校准的二极管)。通过与胶片比较和将检测器旋转180度来验证阵列的相对校准。当用均匀的幻象材料代替探测器板之一时,剂量计对旋转辐射的响应变化不超过0.2%。每日输出校正因子可从均匀圆柱场中的Delta4测量得出。 γ(3%,3 mm)通过率(绝对剂量)在所有9个评估的临床计划中均超过90%,在除一项以外的所有其他方案中均超过96%。平均及格率为97 +/- 2.7%。这些计划在调制系数,音高和计算网格大小方面有所不同。为了获得最佳结果,需要仔细对准体模,最好使用兆伏计算机断层扫描成像。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号