Design and evaluation of a handheld impedance plethysmograph for measuring heart rate variability.

机译:用于测量心率变异性的手持式阻抗体积描记器的设计和评估。

获取原文
获取原文并翻译 | 示例
           

摘要

Heart rate variability (HRV) analysis from 10s ECGs has been shown to be reliable. However, the short examination time warrants a user-friendly system that can be used for ad-hoc examinations without normal preparation, unlike ECG. A handheld device has been developed that can measure ultra-short HRV from impedance plethysmographic recordings of the pulse wave in distal superficial arteries. The prototype device was made user-friendly through a compact, pen-like design and the use of integrated metal electrodes that were especially designed for dry operation. The main signal processing was performed by a digital signal processor, where the discrete heart beats were detected using a correlation algorithm that could adapt to individual pulse wave shapes to account for biological variation. The novel device was evaluated in 20 mainly young volunteers, using 10 s time-correlated ECG recordings as the reference method. Agreement between the two methods in measuring heart rate and root mean square of successive differences in the heart beat interval (RMSSD) was analysed using correlation coefficients (Pearson's R2), mean differences with 95% confidence intervals and 95% limits of agreement, and Bland-Altman plots. The correlation between the two methods was R2 = 1.00 and R2 = 0.99 when heart rate and RMSSD were measured, respectively. The Bland-Altman plots showed suitable agreement between the novel device and standard 10 s ECGs, which was substantiated by 95% limits of agreement of the difference of +/- 0.1 beats min(-1) and approximately +/- 10 ms for heart rate and RMSSD, respectively. Therefore the evaluation showed no significant systematic error of the novel device compared with ECG.
机译:研究表明,来自10s心电图的心率变异性(HRV)分析是可靠的。但是,与ECG不同,检查时间短,保证了用户友好的系统,无需常规准备即可用于即席检查。已经开发了一种手持设备,其可以从远端浅表动脉的脉搏波的阻抗体积描记记录测量超短HRV。通过紧凑的笔状设计以及专为干运行设计的集成金属电极,使原型设备变得用户友好。主要信号处理由数字信号处理器执行,其中使用相关算法检测离散心跳,该算法可以适应各个脉搏波形以解决生物学差异。使用10 s时间相关的ECG记录作为参考方法,对20位主要是年轻志愿者进行了评估。使用相关系数(Pearson's R2),具有95%置信区间和95%一致性限制的均值差异以及Bland分析了两种测量心率和心跳间隔连续差异均方根的方法之间的一致性(Pearson's R2) -奥特曼剧情。当测量心率和RMSSD时,两种方法之间的相关性分别为R2 = 1.00和R2 = 0.99。 Bland-Altman图显示了新型装置与标准10 s心电图之间的适当一致性,这由95%的+/- 0.1次搏动min(-1)和大约+/- 10 ms的心脏差异的一致性极限所证实速率和RMSSD。因此,评估结果表明,与ECG相比,该新型设备没有明显的系统误差。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号