首页> 外文期刊>Medical dosimetry: official journal of the American Association of Medical Dosimetrists >Dosimetric characteristics of the Siemens IGRT carbon fiber tabletop.
【24h】

Dosimetric characteristics of the Siemens IGRT carbon fiber tabletop.

机译:西门子IGRT碳纤维桌面的剂量特性。

获取原文
获取原文并翻译 | 示例
       

摘要

In this work, the dosimetric characteristics of a new commercial carbon fiber treatment table are investigated. The photon beam attenuation properties of the Siemens image-guided radiation therapy (IGRT) tabletop were studied in detail. Two sets of dosimetric measurements were performed. In the first experiment a polystyrene slab phantom was used: the central axis attenuation and the skin-sparing detriment were investigated. In the second experiment, the off-axis treatment table transmission was investigated using a polystyrene cylindrical phantom. Measurements were taken at the isocenter for a 360 degrees rotation of the radiation beam. Our results show that the photon beam attenuation of the Siemens IGRT carbon fiber tabletop varies from a minimum of 2.1% (central axis) to a maximum of 4.6% (120 degrees and 240 degrees beam incidence). The beam entrance dose increases from 82% to 97% of the dose at the depth of maximum for a clinical 6-MV radiation field. The depth of maximum also decreases by 0.4 cm. Despite the wedge cross section of the table the beam attenuation properties of the IGRT tabletop remain constant along the longitudinal direction. American Association of Medical Dosimetrists.
机译:在这项工作中,研究了新型商用碳纤维处理台的剂量特性。详细研究了西门子图像引导放射治疗(IGRT)桌面的光子束衰减特性。进行了两组剂量学测量。在第一个实验中,使用了聚苯乙烯平板模型:研究了中心轴衰减和保留皮肤的损害。在第二个实验中,使用聚苯乙烯圆柱体模型研究了离轴治疗台的传动情况。在等角点对辐射束旋转360度进行测量。我们的结果表明,西门子IGRT碳纤维桌面的光子束衰减范围从最小2.1%(中心轴)到最大4.6%(120度和240度光束入射)不等。对于临床6-MV辐射场,在最大深度处,光束入射剂量从剂量的82%增加到97%。最大深度也减少了0.4厘米。尽管桌子的横截面是楔形的,但IGRT桌面的光束衰减特性仍沿纵向保持恒定。美国医学剂量学协会。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号