Effects of neurosurgical titanium mesh on radiation dose.

机译:神经外科钛网对辐射剂量的影响。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

The purpose of this study was to determine the dosimetric impact of a neurosurgical titanium mesh in patients treated with 6- and 18-MV photon beams. The effects of a 0.4-mm-thick titanium mesh on the dose profile at 3 regions within a solid water phantom were measured using extended dose range-2 (EDR2) film for 6- and 18-MV photon beams. All measurements were performed with the titanium mesh placed at a depth of 1.5 cm in the phantom. Films were exposed immediately above the mesh, immediately below the mesh, and at a depth of 5 cm from the surface of the phantom. The films were scanned using a scanning densitometer. In the region directly above the titanium mesh, there was an increase in dose of 7.1% for 6-MV photons and 4.9% for 18-MV photons. Directly below the titanium mesh, there was an average decrease in dose of 1.5% for 6-MV photons and an increase of 1.0% for 18-MV photons. At 5-cm depth, for 6- and 18-MV photons, there was a decrease in dose of 2.2% and 0.6%, respectively. We concluded that for cranial irradiation with high-energy photons, the dosimetric impact of a 0.4-mm titanium mesh is small and does not require modification in treatment parameters.
机译:这项研究的目的是确定神经外科钛网对6和18 MV光子束治疗的患者的剂量学影响。对于6和18 MV光子束,使用扩展剂量范围2(EDR2)膜测量了0.4毫米厚的钛网对固体水体模中3个区域的剂量分布的影响。所有的测量都是将钛网放置在体模中1.5厘米深处进行的。膜在网眼正上方,网眼正下方和距幻影表面5 cm的深度处曝光。使用扫描密度计扫描膜。在钛网上方的区域中,6-MV光子的剂量增加了7.1%,而18-MV光子的剂量增加了4.9%。在钛网的正下方,6-MV光子的平均剂量减少了1.5%,而18-MV光子的平均剂量增加了1.0%。在5厘米深度处,对于6和18 MV光子,剂量分别减少2.2%和0.6%。我们得出的结论是,对于高能光子的颅骨照射,0.4毫米钛网的剂量学影响很小,不需要修改治疗参数。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号