A non-contact vital sign monitoring system for ambulances using dual-frequency microwave radars.

机译:使用双频微波雷达的救护车非接触式生命体征监测系统。

获取原文
获取原文并翻译 | 示例
       

摘要

We developed a novel non-contact monitoring system to measure the vital signs of casualties inside a moving ambulance. This system was designed to prevent exposure of patients to infectious organisms under biochemical hazard conditions. The system consists of two microwave radars: a 10-GHz respiratory-monitoring radar is positioned 20 cm away from the surface of the isolator. The 24-GHz cardiac-monitoring radar is positioned below the stretcher underneath the isolator. The subject (22.13 +/- 0.99 years) was placed inside the isolator on a stretcher in the simulated ambulance. While the ambulance was in motion at a speed of approximately 10 km/h, the heart rates determined by the cardiac-monitoring radar correlated significantly with those measured by ECG (r = 0.69, p < 0.01), and the respiratory rates derived from the respiratory-monitoring radar correlated with those measured by the respiration curves (r = 0.97, p < 0.0001). The proposed system appears promising for future on-ambulance monitoring of the vital sign of casualties exposed to toxins.
机译:我们开发了一种新颖的非接触式监控系统,用于测量移动中的救护车内人员伤亡的生命体征。该系统旨在防止患者在生化危险条件下接触传染性生物。该系统由两个微波雷达组成:一个10 GHz呼吸监测雷达,距离隔离器表面20厘米。 24 GHz心脏监测雷达位于隔离器下方的担架下方。将受试者(22.13 +/- 0.99岁)放在模拟救护车中担架上的隔离器内。当救护车以大约10 km / h的速度运动时,由心脏监护雷达确定的心率与由ECG测量的心率显着相关(r = 0.69,p <0.01),并且呼吸速率源自呼吸监测雷达与通过呼吸曲线测得的雷达相关(r = 0.97,p <0.0001)。拟议中的系统似乎很有希望用于将来的救护车监测暴露于毒素的人员伤亡的生命体征。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号