Comparison of Kodak EDR2 and Gafchromic EBT film for intensity-modulated radiation therapy dose distribution verification.

机译:比较柯达EDR2和全色EBT胶片进行强度调节的放射治疗剂量分布验证。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

The quantitative dose validation of intensity-modulated radiation therapy (IMRT) plans require 2-dimensional (2D) high-resolution dosimetry systems with uniform response over its sensitive region. The present work deals with clinical use of commercially available self-developing Radio Chromic Film, Gafchromic EBT film, for IMRT dose verification. Dose response curves were generated for the films using a VXR-16 film scanner. The results obtained with EBT films were compared with the results of Kodak extended dose range 2 (EDR2) films. The EBT film had a linear response between the dose range of 0 to 600 cGy. The dose-related characteristics of the EBT film, such as post irradiation color growth with time, film uniformity, and effect of scanning orientation, were studied. There was up to 8.6% increase in the color density between 2 to 40 hours after irradiation. There was a considerable variation, up to 8.5%, in the film uniformity over its sensitive region. The quantitative differences between calculated and measured dose distributions were analyzed using DTA and Gamma index with the tolerance of 3% dose difference and 3-mm distance agreement. The EDR2 films showed consistent results with the calculated dose distributions, whereas the results obtained using EBT were inconsistent. The variation in the film uniformity limits the use of EBT film for conventional large-field IMRT verification. For IMRT of smaller field sizes (4.5 x 4.5 cm), the results obtained with EBT were comparable with results of EDR2 films.
机译:强度调制放射治疗(IMRT)计划的定量剂量验证需要二维(2D)高分辨率剂量测定系统,该系统在其敏感区域上具有一致的响应。本工作涉及用于IMRT剂量验证的可商购的自行开发的放射性铬膜,全色EBT膜的临床应用。使用VXR-16胶片扫描仪生成胶片的剂量响应曲线。将使用EBT胶片获得的结果与柯达扩展剂量范围2(EDR2)胶片的结果进行比较。 EBT膜在0到600 cGy的剂量范围之间具有线性响应。研究了EBT膜的剂量相关特性,例如辐照后随时间的颜色增长,膜的均匀性以及扫描方向的影响。照射后2到40小时之间,颜色密度最多增加8.6%。在其敏感区域上的薄膜均匀性存在很大的差异,最高可达8.5%。使用DTA和Gamma指数分析了计算的剂量分布和测量的剂量分布之间的定量差异,公差为3%剂量差异和3-mm距离一致性。 EDR2胶片显示与计算的剂量分布一致的结果,而使用EBT获得的结果不一致。膜均匀性的变化限制了EBT膜在常规大视野IMRT验证中的使用。对于较小尺寸(4.5 x 4.5厘米)的IMRT,使用EBT获得的结果与EDR2胶片的结果相当。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号