首页> 外文期刊>Medical dosimetry: official journal of the American Association of Medical Dosimetrists >Practical considerations for electron beam small field size dosimetry.
【24h】

Practical considerations for electron beam small field size dosimetry.

机译:电子束小场尺寸剂量学的实际考虑。

获取原文
获取原文并翻译 | 示例
       

摘要

Special care of superficial lesions surrounding critical structures, such as an eye, may require tight margins. When this is the case, small megavoltage electron treatment fields and nonstandard treatment distances become necessary. When the field size is found to be less than the practical range of the electron beam, dosimetric measurements should be performed. This research includes data proving that very small electron fields can be employed for treatment with appropriate beam flatness and penumbra. This is accomplished by first coming down the incident beam to a small field size, then secondly by adding a single lead sheet to the patient's skin surface. The aperture of the sheet is required to be greater than 2 x 2 cm2 in size, and must be cut properly to adequately confine the treatment area.
机译:对关键结构(例如眼睛)周围的浅表病变进行特殊护理可能需要紧紧的边缘。在这种情况下,小的兆电压电子处理场和非标准的处理距离变得必要。当发现场大小小于电子束的实际范围时,应进行剂量测定。这项研究包括的数据证明,很小的电子场可用于以适当的光束平坦度和半影进行处理。这是通过首先将入射光束降到较小的视野大小,然后将单个铅片添加到患者的皮肤表面来实现的。片材的孔径要求大于2 x 2 cm2,并且必须适当切割以充分限制治疗区域。