A novel multislice B 1-mapping method dubbed dual refocusing echo acquisition mode is proposed, able to cover the whole transmit coil volume in only one second, which is more than an order of magnitude faster than established approaches. The dual refocusing echo acquisition mode technique employs a stimulated echo acquisition mode (STEAM) preparation sequence followed by a tailored single-shot gradient echo sequence, measuring simultaneously the stimulated echo and the free induction decay as gradient-recalled echoes, and determining the actual flip angle of the STEAM preparation radiofrequency pulses from the ratio of the two measured signals. Due to an elaborated timing scheme, the method is insensitive against susceptibility/chemical shift effects and can deliver a B 0 phase map and a transceive phase map for free. The approach has only a weak T 1 and T 2 dependence and moreover, causes only a low specific absorption rate (SAR) burden. The accuracy of the method with respect to systematic and statistical errors is investigated both, theoretically and in experiments on phantoms. In addition, the performance of the approach is demonstrated in vivo in B 1-mapping and radiofrequency shimming experiments on the abdomen, the legs, and the head on an eight-channel parallel transmit 3 T MRI system.
机译:提出了一种新颖的多层B 1映射方法,称为双重重聚焦回波获取模式,该方法能够在1秒钟内覆盖整个发射线圈的体积,比已建立的方法快一个数量级。双重重聚焦回波采集模式技术采用刺激回波采集模式(STEAM)准备序列,然后采用定制的单次梯度回波序列,同时测量激发回波和自由感应衰减作为梯度回波,并确定实际翻转从两个测量信号之比得出STEAM准备射频脉冲的角度。由于精心设计的时序方案,该方法对磁化率/化学位移效应不敏感,并且可以免费提供B 0相图和收发相图。该方法仅具有弱的T 1和T 2依赖性,而且仅引起较低的比吸收率(SAR)负担。理论上和模型实验中都研究了该方法在系统误差和统计误差方面的准确性。此外,该方法的性能已在八通道并行传输3 T MRI系统的腹部,腿部和头部的B 1映射和射频匀场实验中得到了证实。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号