Hyperpolarized 129Xe gas lung MRI-SNR and T 2 * comparisons at 1.5 T and 3 T

机译:1.5 T和3 T时超极化129Xe气肺MRI-SNR和T 2 *比较

获取原文
获取原文并翻译 | 示例
           

摘要

In this study, the signal-to-noise ratio of hyperpolarized 129Xe human lung magnetic resonance imaging was compared at 1.5 T and 3 T. Experiments were performed at both B 0 fields with quadrature double Helmholtz transmit-receive chest coils of the same geometry with the same subject loads. Differences in sensitivity between the two field strengths were assessed from the signal-to-noise ratio of multi-slice 2D 129Xe ventilation lung images obtained at the two field strengths with a spatial resolution of 15 mm × 4 mm × 4 mm. There was a systematically higher signal-to-noise ratio observed at 3 T than at 1.5 T by a factor of 1.25. Mean image signal-to-noise ratio was in the range 27-44 at 1.5 T and 36-51 at 3 T. T 2* of 129Xe gas in the partially inflated lungs was measured to be 25 ms and 18 ms at 1.5 T and 3 T, respectively. T 2* of 129Xe gas in fully inflated lungs was measured to be 52 ms and 24 ms at 1.5 T and 3 T, respectively.
机译:在这项研究中,在1.5 T和3 T下对超极化129Xe人肺磁共振成像的信噪比进行了比较。在两个B 0场均进行了实验,两个正交的Helmholtz收发器具有相同的几何形状,相同的主题负荷。根据在两个空间强度为15 mm×4 mm×4 mm的两个场强下获得的多层2D 129Xe通风肺图像的信噪比,评估了两个场强之间的灵敏度差异。在3 T时观察到的系统信噪比比1.5 T时系统地高1.25倍。平均图像信噪比在1.5 T时为27-44,在3 T时为36-51。在局部充胀的肺中,129Xe气体的T 2 *在1.5 T和25 T时分别为25 ms和18 ms。 3 T,分别。完全膨胀的肺中129Xe气体的T 2 *在1.5 T和3 T时分别为52 ms和24 ms。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号