PURPOSE: To report a potentially serious complication resulting from a faulty guide wire during central venous catheterization. CLINICAL FEATURES: We report breakage of a guide wire with potential severe implications for a 47-yr-old critical care patient. The tip of the J-wire became lodged in the internal jugular vein, and required fluoroscopic guidance for removal, which occurred without complication. CONCLUSION: Breakage of a guide wire during central venous catheter insertion has been very rarely reported. However, inherent faults in design or in the manufacturing process of the guide wire could lead to this rare complication. A simple bedside test is proposed to detect breakage in the core section of the guide wire.
机译:目的:报告在中央静脉导管插入过程中由于导线故障而可能引起的严重并发症。临床特征:我们报告导丝断裂,对一名47岁的重症监护患者可能造成严重的影响。 J线的尖端卡在了颈内静脉中,需要通过荧光镜检查切除,这没有并发症。结论:极少有报道在中央静脉导管插入过程中导丝断裂。但是,导丝的设计或制造过程中固有的故障可能导致这种罕见的复杂情况。提出了一种简单的床头测试,以检测导丝芯部的断裂。