An unusual failure of a laryngoscope handle.

机译:喉镜手柄的异常故障。

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

To the Editor:A 22-year-old male with hyperhidrosis presented for general anesthesia for a scheduled bilateral video-assisted thoracoscopy and sympathectomy. During the laryngoscopy, the handle of the Rusch (Sulzbach, Germany) laryngoscope blade separated at the weld line (a 3 Macintosh blade was used to place a double lumen endotracheal tube) (Fig. 1). As a result, the patient's trachea was intubated with a backup laryngoscope handle and 3 Macintosh blade, both of which had been previously inspected. In retrospect, the failure of the weld line could have been caused by a number of factors; such as, a hairline fracture, subtle damage if the handle was previously dropped, repeated disinfection with germicidal disposable wipes (PDI Sani-cloth~(TM)), or repeated cleaning with soap and water. This particular laryngoscope handle had been in service for approximately three years.
机译:致编辑:一名22岁多汗男性因全麻进行了定期双侧胸腔镜和交感神经切除术。喉镜检查期间,Rusch(德国苏尔茨巴赫)喉镜刀片的手柄在焊缝处分开(3 Macintosh刀片用于放置双腔气管导管)(图1)。结果,用备用喉镜手柄和3台Macintosh刀片向患者的气管插管,这两项都是先前检查过的。回想起来,焊接线的故障可能是由多种因素引起的;例如发际线断裂,如果先前掉落手柄会造成细微的损坏,用杀菌的一次性湿巾(PDI Sani-cloth〜)重复消毒或用肥皂和水反复清洗。这种特殊的喉镜手柄已经使用了大约三年。