Failure of the Airway Scope to reach the larynx.

机译:气道镜无法到达喉头。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

The Airway Scope (AWS) is a useful and possibly promising device for orotracheal intubation in a patient with a difficult airway, because it provides an excellent view of the glottis without requiring alignment of the oral, pharyngeal, and laryngeal axes. We, however, encountered a case in which the INTLOCK, the blade of the AWS, could not proceed beneath the epiglottis, and consequently, the AWS failed to intu-bate the patient's trachea, probably because the distance from the mouth to the larynx of the patient was longer than the designed length of the INTLOCK.
机译:气道镜(AWS)是一种在气道困难的患者中进行气管插管的有用且可能很有前途的设备,因为它可提供声门的绝佳视野,而无需对齐口腔,咽喉和喉轴。但是,我们遇到了这样的情况:AWS的刀片INTLOCK无法在会厌下方行进,因此,AWS无法将患者的气管植入体内,这可能是因为从嘴到喉咙的距离患者比INTLOCK的设计长度长。