首页> 外文期刊>Canadian journal of anesthesia: Journal canadien d'anesthesie >Production of compound A under low-flow anesthesia is affected by type of anesthetic machine.
【24h】

Production of compound A under low-flow anesthesia is affected by type of anesthetic machine.

机译:低流量麻醉下化合物A的生产受麻醉机类型的影响。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

PURPOSE: The purpose was to compare the concentrations of compound A in inspired gas breathed by patients produced by different types of anesthetic machines under prolonged sevoflurane low-flow anesthesia. METHODS: The anesthetic machines tested were Excel 210 SE (Datex-Ohmeda, Louisville, CO), Cicero (Drager, Lubeck, Germany), and AS/3 ADU (Datex-Ohmeda, Louisville, CO). Anesthesia expected to last more than four hours was maintained with 2.0% sevoflurane and nitrous oxide (0.5 L x min(-1))/oxygen (0.5 L x min(-1)). The concentrations of compound A, obtained from the inspiratory limb of the circle system, were measured using a gas chromatograph. RESULTS: When Excel and Cicero were used, concentrations of compound A increased steadily from the baseline values to 28 and 29 (mean) ppm, respectively, at two hours after exposure to sevoflurane and became constant. There was no significant difference between the concentrations of compound A produced by these anesthetic machines. In contrast, the new anesthetic machine AS/3 was associated with lower concentrations of compound A (6 ppm at one hour, P <0.05 compared with Excel and Cicero), and the concentration did not change significantly thereafter. CONCLUSION: In spite of the use of a conventional carbon dioxide (CO2) absorbent with strong bases, the anesthetic machine AS/3 with a small volume of canister/soda lime (900 ml/700 ml) produced lower concentrations of compound A than those produced by the other machines.
机译:目的:目的是比较长期在七氟醚低流量麻醉下不同类型麻醉机产生的患者所吸入的呼吸气体中化合物A的浓度。方法:所测试的麻醉机为Excel 210 SE(达克斯-欧米达,路易斯维尔,科罗拉多州),西塞罗(德拉格,吕贝克,德国)和AS / 3 ADU(达克斯-奥默达,路易斯维尔,科罗拉多州)。使用2.0%的七氟醚和一氧化二氮(0.5 L x min(-1))/氧气(0.5 L x min(-1))维持预期持续超过4小时的麻醉。使用气相色谱仪测量从圆形系统的吸气肢获得的化合物A的浓度。结果:当使用Excel和Cicero时,在暴露于七氟醚的两个小时后,化合物A的浓度分别从基准值稳定地增加到28和29(平均)ppm,并保持恒定。这些麻醉机产生的化合物A的浓度之间没有显着差异。相反,新的麻醉机AS / 3与较低浓度的化合物A相关联(一小时为6 ppm,与Excel和Cicero相比,P <0.05),并且此后浓度没有明显变化。结论:尽管使用具有强碱的常规二氧化碳(CO2)吸收剂,麻醉机AS / 3的罐/苏打石灰量较小(900 ml / 700 ml),其化合物A的浓度低于那些由其他机器生产。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号