首页> 外文期刊>Canadian journal of anesthesia: Journal canadien d'anesthesie >Failure of a water-bath design intravenous fluid warmer.
【24h】

Failure of a water-bath design intravenous fluid warmer.

机译:水浴设计失败的静脉输液器。

获取原文
获取原文并翻译 | 示例

摘要

We write to highlight the failure of an intravenous fluid warming device, HOTLINE~R HL-290 Blood and Fluid Warmer (Smiths Medical International, Kent, UK), which could have resulted in harm to the patient. Fluid warmer failures are becoming more prevalent, as evidenced by the United Kingdom Medicines and Healthcare Products Regulatory Agency (MHRA) issuing a recall of all Level 1 SIMS fluid warmer sets.The fluid warmer in our case is designed differently. Warm water is pumped into a counter-current system around an inner sterile infusion lumen. The warming fluid is kept in a non-sterile water bath, emphasizing the need for extreme care with all water-bath fluid warmers.
机译:我们着重强调静脉输液加温装置HOTLINE〜R HL-290血液和液体加温器(英国肯特郡史密斯医学国际公司)的故障,该故障可能会对患者造成伤害。英国药品和保健产品监管局(MHRA)发出召回所有1级SIMS液体加热器套件的证据表明,液体加热器故障变得越来越普遍。在我们的案例中,液体加热器的设计有所不同。将温水泵入内部无菌输液腔周围的逆流系统中。加热液被保存在非无菌水浴中,强调所有水浴液加热器都需要格外小心。