首页> 外文期刊>Canadian journal of anesthesia: Journal canadien d'anesthesie >Transcutaneously measured near-infrared spectroscopic liver tissue oxygenation does not correlate with hepatic venous oxygenation in children: (La mesure transcutanee de l'oxygenation du tissu hepatique par spectroscopie proche infrarouge ne correspo
【24h】

Transcutaneously measured near-infrared spectroscopic liver tissue oxygenation does not correlate with hepatic venous oxygenation in children: (La mesure transcutanee de l'oxygenation du tissu hepatique par spectroscopie proche infrarouge ne correspo

机译:经皮测量的近红外光谱肝组织氧合与儿童肝静脉氧合无关:

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

PURPOSE: To compare transcutaneous near-infrared spectroscopic (NIRS) measured liver tissue oxygenation with hepatic vein oxygen saturation (SvhO(2)) in children undergoing cardiac catheterization. METHODS: A NIRS optode (containing an emitter and a receiver of near-infrared light) was placed directly below the right costal arch above the palpable liver in 40 children aged 0.02 to 7.28 yr (median: 1.8 yr). Spatially resolved spectroscopic measured tissue oxygenation index (TOI) was recorded using the NIRO-300(TM). Paired blood samples from the hepatic vein were taken under radiological control for determination of SvhO(2) in a co-oxymeter. TOI values were compared with hepatic vein oxygenation, with simultaneously obtained arterial oxygen saturation (SaO(2)), inferior vena cava SO(2) and hemoglobin concentration using simple linear and multi-regression analysis. RESULTS: TOI values ranged from 35% to 73% (58.6 +/- 8.4%); SvhO(2) from 32% to 80% (58.4 +/- 14.4%), and arterial SO(2) from 54% to 100% (90.0 +/- 11.4%). TOI and hepatic vein oxygen saturation failed to correlate (r = 0.052/P = 0.752). A regression model containing arterial saturation ( triangle up R(2) = 0.177) and the ratio of pulmonary to systemic resistance ( triangle up R(2) = 0.095) explained 27.3% of the observed variance in TOI. In this model, hepatic vein oxygen saturation was no longer significant; explaining only 3.4% of the variance. No other variable retained a significant association. CONCLUSION: Transcutaneously measured NIRS tissue oxygenation with an optode placed over the palpable liver does not correlate with SvhO(2). The value is dominated by non-hepatic variables such as arterial saturation and vascular resistances.
机译:目的:为了比较经心导管检查的儿童经皮近红外光谱(NIRS)测量的肝组织氧合与肝静脉血氧饱和度(SvhO(2))。方法:将40个0.02至7.28岁(中位数:1.8岁)儿童的NIRS光电二极管(包含近红外光发射器和接收器)放置在可触及肝脏上方右肋弓下方。使用NIRO-300 TM记录空间分辨光谱测量的组织氧合指数(TOI)。在放射学控制下从肝静脉中抽取成对的血液样本,在血氧仪中测定SvhO(2)。使用简单的线性和多元回归分析比较TOI值与肝静脉氧合,同时获得动脉血氧饱和度(SaO(2)),下腔静脉SO(2)和血红蛋白浓度。结果:TOI值范围从35%到73%(58.6 +/- 8.4%); SvhO(2)从32%增至80%(58.4 +/- 14.4%),动脉SO(2)从54%增至100%(90.0 +/- 11.4%)。 TOI和肝静脉血氧饱和度不相关(r = 0.052 / P = 0.752)。包含动脉饱和度(三角向上R(2)= 0.177)和肺部与全身阻力的比率(三角向上R(2)= 0.095)的回归模型解释了TOI中观察到的变化的27.3%。在该模型中,肝静脉血氧饱和度不再显着。只能解释方差的3.4%。没有其他变量保留显着的关联。结论:经皮测得的NIRS组织氧合,并在可触及的肝脏上方放置一光极与SvhO(2)不相关。该值主要受非肝脏变量(如动脉饱和度和血管阻力)的影响。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号