Inferior stability of a biodegradable cement plug : 122 total hip replacements randomized to degradable or non-degradable cement restrictor.

机译:可生物降解水泥塞的稳定性较差:122 例髋关节置换术随机分配到可降解或不可降解的水泥限制器组。

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

Introduction: Modern cementing technique demands high intramedullar cement pressure to create an optimum fixed cement support of the femoral stem. Pressure resistant, stable closure of the canal therefore is absolutely necessary. Biodegradable cement restrictors, if corresponding to above mentioned attributes, could be of immense advantage in case of potential revision surgery. Material and methods: A prospective, randomized trial was performed on a consecutive series of 130 patients who underwent primary cemented total hip arthroplasty due to hip arthrosis to compare a degradable cement restrictor and a non-degradable cement restrictor in their ability to resist distal migration during stem insertion. Results: The median cement plug length measured 27 mm (range -12 to 126 mm, 95 confidence interval (CI) 20-33 mm) in the biodegradable restrictor group and 15 mm (range 0-61 mm, 95 CI 12-18 mm) in the non-degradable restrictor group (P=0.003). A significant effect of the relationship between the difference of restrictor size and the diameter of the femoral canal on length of cement distal of the tip of the stem in between the two groups was evident (P=0.031). Conclusion: The results indicate insufficient intramedullary plug fixation of the degradable restrictor probably due to the elastic material properties which also may lead to inferior precision in restrictor size choice.
机译:简介:现代骨水泥技术需要高髓内水泥压力,以形成股骨柄的最佳固定水泥支撑。因此,耐压、稳定地关闭运河是绝对必要的。可生物降解的水泥限制剂,如果符合上述属性,在潜在的翻修手术中可能具有巨大的优势。材料和方法:对 130 例因髋关节病而接受原发性骨水泥全髋关节置换术的患者进行了一项前瞻性随机试验,以比较可降解骨水泥限制剂和不可降解骨水泥限制剂在茎插入过程中抵抗远端迁移的能力。结果:可生物降解限制器组的中位水泥塞长度为 27 mm(范围 -12 至 126 mm,95% 置信区间 (CI) 20-33 mm),不可降解限制器组为 15 mm(范围 0-61 mm,95% CI 12-18 mm)(P=0.003)。两组限制器大小与股骨管直径的差异对骨干尖端远端水泥长度的影响显著(P=0.031)。结论:结果表明,可降解限流器髓内塞固定不充分,可能是由于弹性材料特性所致,这也可能导致限流器尺寸选择精度较低。

著录项

获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号