首页> 外文期刊>Veterinary Radiology & Ultrasound >A NOVEL, REMOVABLE, CERROBEND, BEAM-BLOCKING DEVICE FOR RADIATION THERAPY OF THE CANINE HEAD AND NECK: PILOT STUDY
【24h】

A NOVEL, REMOVABLE, CERROBEND, BEAM-BLOCKING DEVICE FOR RADIATION THERAPY OF THE CANINE HEAD AND NECK: PILOT STUDY

机译:一种用于犬头颈部放射治疗的新型、可拆卸、CERROBEND、阻束装置:初步研究

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

Radiation therapy of the head and neck can result in mucositis and other acute affects in the oral cavity. This prospective pilot study evaluated a novel, intraoral, beam-blocking device for use during imaging and therapeutic procedures. The beam-blocking device was made from a metal alloy inserted into a coated frozen dessert mold (Popsicle (R) Mold, Cost Plus World Market, Oakland, CA). The device was designed so that it could be inserted into an outer shell, which in turn allowed it to be placed or removed depending on the need due to beam configuration. A Farmer type ionization chamber and virtual water phantom were used to assess effects of field size on transmission. Six large breed cadaver dogs, donated by the owner after death, were recruited for the study. Delivered dose at the dorsal and ventral surfaces of the device, with and without the alloy block in place, were measured using radiochromic film. It was determined that transmission was field size dependent with larger field sizes leading to decreased attenuation of the beam, likely secondary to scatter. The mean and median transmission on the ventral surface without the beam-blocking device was 0.94 range 0.94-0.96. The mean and median transmission with the beam-blocking device was 0.52 range 0.50-0.57. The mean and median increase in dose due to backscatter on the dorsal surface of the beam-blocking device was 0.04 range 0.02-0.04. Findings indicated that this novel device can help attenuate radiation dose ventral to the block in dogs, with minimal backscatter. (C) 2016 American College of Veterinary Radiology.
机译:头部和颈部的放射治疗可导致口腔粘膜炎和其他急性影响。这项前瞻性试点研究评估了一种用于成像和治疗程序的新型口内阻束装置。光束阻挡装置由插入涂层冷冻甜点模具(冰棒(R)模具,成本加成世界市场,加利福尼亚州奥克兰)的金属合金制成。该装置的设计使其可以插入外壳中,这反过来又允许根据梁配置的需要放置或移除它。使用Farmer型电离室和虚拟水模型来评估田间大小对传输的影响。六只大型尸体犬被招募,由主人在死后捐赠,用于这项研究。使用放射致变色胶片测量设备背面和腹面的递送剂量,无论是否放置合金块。确定透射与场大小有关,较大的场大小导致光束衰减减少,可能继发于散射。在没有光束阻挡装置的情况下,腹面的平均和中位透射率为0.94 [范围0.94-0.96]。使用光束阻挡装置的平均和中值透射率为0.52 [范围0.50-0.57]。由于束流阻挡装置背面的反向散射导致的剂量平均和中位增加为 0.04 [范围 0.02-0.04]。研究结果表明,这种新型装置可以帮助减弱狗腹侧对块的辐射剂量,同时将反向散射降至最低。(C) 2016 年美国兽医放射学会。

著录项

获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号