首页> 外文期刊>Foot and ankle international >A mechanical equinometer to measure the range of motion of the ankle joint: interobserver and intraobserver reliability.
【24h】

A mechanical equinometer to measure the range of motion of the ankle joint: interobserver and intraobserver reliability.

机译:一种机械等值仪,用于测量踝关节的运动范围:观察者之间和观察者内部的可靠性。

获取原文
获取原文并翻译 | 示例
           

摘要

BACKGROUND: Clinical measurement of passive dorsiflexion of the ankle joint is essential for the diagnosis of various pathologic conditions of the foot and ankle but is of unreliable precision with high interobserver variability in nonweightbearing tests. This work was designed to develop and test a precise, standardized, and reliable technique for measurement of passive and active ankle range of motion. METHODS: The proposed measurement tool is composed of two mobile parallelograms, one attached to the tibia, the second one to the plantar surface of the foot. The parallelograms are connected with a hinge with an angular scale to measure the angle between the foot and tibia. RESULTS: Interobserver correlation between clinical measurements for maximal passive foot dorsiflexion were 0.03 with knee extension and 0.38 with knee flexion, while for measurements with the proposed tool they reached 0.89 and 0.97, respectively, with a mean measurement error of 0.9 degrees. Intraobserver correlations reached values of r = 0.98 and 0.99. CONCLUSIONS: The proposed tool allows measurement of the ankle range of motion with very high precision and reproducibility far superior to clinical measurements. CLINICAL RELEVANCE: Precise measurement of ankle range of motion is clinically challenging. With the use of the proposed tool, measurement precision and reliability are decisively improved.
机译:背景:踝关节被动背屈的临床测量对于诊断足部和踝关节的各种病理状况至关重要,但在非负重测试中,其准确性不可靠且观察者间差异很大。这项工作旨在开发和测试一种精确,标准化和可靠的技术,用于测量被动和主动脚踝的运动范围。方法:建议的测量工具由两个可移动的平行四边形组成,一个固定在胫骨上,第二个固定在足底表面上。平行四边形与带有角度刻度的铰链连接,以测量脚和胫骨之间的角度。结果:观察到的最大被动脚背屈的临床测量之间的观察者相关性是,膝关节伸展时为0.03,膝关节屈曲时为0.38,而使用建议的工具进行的测量分别为0.89和0.97,平均测量误差为0.9度。观察者内相关性达到r = 0.98和0.99的值。结论:所提出的工具允许以非常高的精度和可再现性测量踝关节的运动范围,远优于临床测量。临床相关性:精确测量踝关节运动范围在临床上具有挑战性。通过使用所提出的工具,可以显着提高测量精度和可靠性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号