OBJECTIVES: Wear of attachments leads to a loss of retention and potentially reduces the function of complete dentures. This study evaluated the retention force changes of different prefabricated attachment systems for implant-supported overdentures to estimate the wear constancy and applicability in clinical practice. METHODS: Four prefabricated attachment systems were tested [Group SG: retentive ball attachment (Straumann, Switzerland) with gold matrix, Group ST: retentive ball attachment (Straumann, Switzerland) with titanium spring matrix, Group IB: UNOR i-Ball with Ecco matrix (UNOR, Switzerland) and Group IMZ: IMZ-TwinPlus ball attachment with gold matrix (DENTSPLY Friadent, Germany)]. Ten samples of each system were subjected to 10,000 insertion-separation cycles. RESULTS: Results showed that all types of attachments showed wear, which led to a loss of retention force after an initial increase at the beginning of the wear simulation. Attachments with a plastic retention insert or gold matrices underwent the smallest changes in retention force. The titanium spring system showed the largest changes in retention force and a greater variation between the different cycles and specimen. This behaviour is probably caused by a large fitting tolerance of the titanium spring. CONCLUSIONS: Attachment systems which possess a male and female component of different material composition are preferable. They show smaller changes in the retention force. For retention force increase and wear compensation, an attachment system should be adjustable.
机译:目的:附件的磨损会导致固位丧失,并可能降低全口义齿的功能。这项研究评估了植入物支持的覆盖义齿的不同预制附着系统的固位力变化,以估计磨损稳定性和在临床实践中的适用性。方法:测试了四个预制的附着系统[SG组:带金基体的保持球附着(瑞士Straumann),ST组:钛合金基体的保持球附着(Straumann,瑞士),IB组:UNOR i-Ball与Ecco矩阵(UNOR,瑞士)和IMZ组:IMZ-TwinPlus带有金矩阵的球附件(DENTSPLY Friadent,德国)]。每个系统的十个样品进行10,000次插入-分离循环。结果:结果表明,所有类型的附件均显示磨损,这在磨损模拟开始时初始增加后导致保持力的损失。带有塑料固定插件或金基的附件​​的固定力变化最小。钛弹簧系统显示出最大的保持力变化,并且不同循环和试样之间的变化更大。此行为可能是由钛弹簧的较大装配公差引起的。结论:具有不同材料组成的公母组件的连接系统是优选的。它们显示出保持力的较小变化。为了增加保持力并进行磨损补偿,附件系统应该是可调节的。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号