I read the recent publication by Johnson et al. In this work, the time domain and spectral domain optical coherence tomography systems were compared. Johnson et al. concluded that "significant differences exist between RNFL measurements obtained from the TD and SD OCT systems used in this study." There are some concerns about this work. First, the small number of tested eyes in the study might be a weakness. Second, there is no information on the quality control of the instrument for measurement in both systems. A common cause of aberration in laboratory diagnostic tests is the error of the analytical tool. In addition, pre-analytical factors, such as the preparation of patients as well as the practitioner's experience, should be a concern.
机译:我读了Johnson等人的最新出版物。在这项工作中,时域和光谱域光学相干断层扫描系统进行了比较。约翰逊等。得出结论,“从本研究中使用的TD和SD OCT系统获得的RNFL测量值之间存在显着差异。”对这项工作有一些担忧。首先,该研究中少数经过测试的眼睛可能是一个弱点。其次,在两个系统中都没有关于用于测量的仪器的质量控制的信息。实验室诊断测试中出现畸变的常见原因是分析工具的错误。另外,分析前的因素,例如患者的准备以及从业者的经验,应该引起关注。