Comparison of high temperature and high Reynolds number water flows between PTB and NMIJ

机译:PTB和NMIJ之间高温和高雷诺数水流的比较

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Bi-comparison for high temperature and high Reynolds number water flows was performed between Physikalische-Technische Bundesanstalt (PTB) and National Metrology Institute of Japan (NMIJ) using differential pressure meters. The flowrate range considered herein was from 250 m(3)/h to 740 m(3)/h and the temperature range was from 20 degrees C to 80 degrees C. The Reynolds number based on the pipe diameter ranges from 4.4x10(5) to 3.6x10(6). The expanded uncertainty of volumetric flowrate with a coverage factor k=2 is 0.040% for PTB and 0.070% for NMIJ. The differential pressure meters as the transfer meter were a throat-tapped flow nozzle and an orifice. The results for both flow meters were in good agreement between the two labs. The difference in the average discharge coefficient at the same Reynolds number ranged from -0.05% to 0.03% for the flow nozzle and from -0.01% to 0.05% for the orifice. The results of the comparison indicate a high level of consistency between PTB and NMIJ for high temperature and high Reynolds number water flows.
机译:使用压差计在Physikalische-Technische Bundesanstalt(PTB)和日本国家计量学会(NMIJ)之间进行了高温和高雷诺数水流量的双比较。本文考虑的流量范围为250 m(3)/ h至740 m(3)/ h,温度范围为20摄氏度至80摄氏度。基于管径的雷诺数范围为4.4x10(5) )至3.6x10(6)。覆盖因子k = 2时体积流量的扩展不确定性对于PTB为0.040%,对于NMIJ为0.070%。作为传输仪表的压差仪表是一个喉头式流量喷嘴和一个节流孔。这两个流量计的结果在两个实验室之间非常吻合。对于相同的雷诺数,流量喷嘴的平均排放系数之差在-0.05%至0.03%的范围内,而对于孔板,则在-0.01%至0.05%的范围内。比较结果表明,PTB和NMIJ在高温和高雷诺数水流量下具有高度一致性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号