首页> 外文期刊>Flow Measurement and Instrumentation >A new calibration facility for water flowrate at high Reynolds number
【24h】

A new calibration facility for water flowrate at high Reynolds number

机译:高雷诺数水流量的新校准设备

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

A new test facility has been constructed for the National Metrology Institute of Japan (NMIJ) and the National Institute of Advanced Industrial Science and Technology (AIST) for calibration of feedwater flowmeters used in nuclear power stations at Reynolds numbers of up to 18 million. This very large Reynolds number is achieved in a 600 mm pipe at a flowrate of 3.33 m{sup}3/s (12,000 m{sup}3/h) and a water temperature of 70℃. This calibration facility consists of a circulation loop with four pumps and four reference flowmeter sets, a prover system, a heating and cooling unit, and other components. The expanded uncertainty of this facility is 0.077%. The present paper describes, in detail, the new facility, the calibration method of the reference flowmeter, experiments for flow field, uncertainty estimation, and the results of an example calibration.
机译:为日本国家计量学会(NMIJ)和国家先进工业科学技术研究院(AIST)建造了一个新的测试设施,用于校准雷诺数高达1800万的核电站中使用的给水流量计。在600 mm的管道中,流速为3.33 m {sup} 3 / s(12,000 m {sup} 3 / h),水温为70℃时,可实现非常大的雷诺数。该校准设备包括一个带有四个泵和四个参考流量计的循环回路,一个检验仪系统,一个加热和冷却单元以及其他组件。此工具的扩展不确定性为0.077%。本文详细介绍了新设备,参考流量计的校准方法,流场实验,不确定性估计以及示例校准的结果。

著录项