摘要

A hot-film anemometer (HFA) calibrator for water was developed that gives flat velocity distribution and low turbulence intensity to liquid flow from a contraction nozzle. The velocity distribution was confirmed using a Pitot tube. The turbulence intensity was investigated by laser Doppler velocimetry (LDV). The values observed by the HFA using a calibrator agreed with the values observed by LDV. The values of computational fluid dynamics (CFD) calculated using the standard k - s model agreed with the values observed by the HFA and LDV. The relationship between the velocity and volumetric flow rate of the liquid was examined by LDV and CFD to investigate the applicability of the calibrator to higher liquid velocity. As a result, the calibrator was found to offer simple operation, correct calibration, and applicability to a wide velocity range.
机译:开发了一种用于水的热膜风速计(HFA)校准器,该校准器对来自收缩喷嘴的液体流提供平坦的速度分布和低的湍流强度。使用皮托管确定速度分布。用激光多普勒测速仪(LDV)研究了湍流强度。 HFA使用校准器观察到的值与LDV观察到的值一致。使用标准k-s模型计算的计算流体动力学(CFD)值与HFA和LDV观察到的值一致。通过LDV和CFD检查液体的速度与体积流量之间的关系,以研究校准器对更高液体速度的适用性。结果,发现该校准器提供了简单的操作,正确的校准以及在宽速度范围内的适用性。