In this study, two methods were developed for the sequential determination of As and Sb, Bi and Pb using fast sequential continuous flow hydride generation atomic absorption spectrometry. Limits of detection obtained for As and Sb were 0.15 and 0.16 mug L-1 and for Bi and Pb were 0.60 and 0.77 mug L-1, respectively. Arsenic, Sb, and Bi were free from the influence of possible interferents in the levels investigated, but Pb showed a significant analytical signal decrease when Co, Cu, Fe, Ni, and HNO3 were presented in the solutions. The methods were applied to the analysis of water from the Tiete River (Sao Paulo State, Brazil), sea water from Praia Grande-Ubatuba (Sao Paulo State), chicken feed, trace elements in water (NIST 1643e), and trace metals in drinking water (HPS TMDW). The obtained recoveries varied from 88 (for As) to 116 % (for Bi). copyright Springer Science+Business Media New York 2012.
机译:在这项研究中,开发了两种使用快速连续连续氢化物发生原子吸收光谱法连续测定As和Sb,Bi和Pb的方法。 As和Sb的检出限分别为0.15和0.16马克杯L -1 ,Bi和Pb的检出限分别为0.60和0.77马克杯L -1 。在所研究的水平上,砷,Sb和Bi不受可能的干扰物的影响,但是当Co,Cu,Fe,Ni和HNO 3 中存在时,Pb显示出明显的分析信号降低。解决方案。该方法用于分析铁特河(巴西圣保罗州)的水,普拉亚格兰德-乌巴图巴(圣保罗州)的海水,鸡饲料,水中的痕量元素(NIST 1643e)和水中的微量金属。饮用水(HPS TMDW)。获得的回收率从88(对于As)到116%(对于Bi)不等。版权所有Springer Science + Business Media纽约,2012年。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号