Experimental contamination assessment of a novel closed ultravitrification device.

机译:新型封闭式超玻璃化装置的实验污染评估。

获取原文
获取原文并翻译 | 示例
           

摘要

OBJECTIVE: To evaluate the safety of a new ultravitrification closed device. DESIGN: Ultravitrification research. SETTING: Private assisted reproduction center. ANIMAL(S): Microorganisms (bacteria). INTERVENTION(S): A styrofoam container holding 1,000 mL of liquid nitrogen (LN2) was contaminated with Pseudomonas aeruginosa and Escherichia coli. Forty closed devices (Ultravit) and 20 open devices (Cryotop) loading approximately 0.5 muL of antibiotic-free medium were plunged into this contaminated LN2 for 5-10 seconds and then inoculated into selective culture dishes. Colony-forming units were analyzed and counted after an overnight incubation at 37 degrees C. MAIN OUTCOME MEASURE(S): Detection of micro-organisms in different devices after incubation. RESULT(S): There was no contamination in any of the closed devices, whereas in 45% of open devices these bacteria were present. CONCLUSION(S): With this study we demonstrated, in an experimental model using contaminated LN2, that this new closed device is a safe system that does not allow cell contact with LN2, avoiding cell contamination.
机译:目的:评估新型超玻璃化封闭设备的安全性。设计:超玻璃化研究。地点:私人辅助生殖中心。动物:微生物(细菌)。干预:装有1,000 mL液氮(LN2)的聚苯乙烯泡沫塑料容器被铜绿假单胞菌和大肠杆菌污染。将40个封闭装置(Ultravit)和20个开放装置(Cryotop)加载约0.5μL不含抗生素的培养基,放入受污染的LN2中5-10秒,然后接种到选择性培养皿中。在37°C过夜孵育后,分析并计数菌落形成单位。主要观察指标:孵育后检测不同设备中的微生物。结果:在任何封闭的设备中都没有污染,而在开放式设备中有45%的细菌存在。结论:通过这项研究,我们在使用受污染的LN2的实验模型中证明了这种新的封闭装置是一种安全的系统,不允许细胞与LN2接触,从而避免了细胞污染。