Repeatability of intra-ocular pressure and central corneal thickness measurements provided by a non-contact method of tonometry and pachymetry.

机译:通过非接触式眼压计和测厚法提供的眼内压和角膜中央厚度测量的可重复性。

获取原文
获取原文并翻译 | 示例
           

摘要

BACKGROUND: To examine the repeatability of intra-ocular pressure (IOP) and central corneal thickness (CCT) measurements provided by the Tono-Pachymeter NT530P (Nidek, Japan), and to assess agreement of the device with Goldmann applanation tonometry (GAT) and ultrasound pachymetry (UP). METHODS: Sixty-two normal eyes of 62 subjects were included in the study. Three sessions of measurements, each consisting of three sets of measurements, were completed with each device. Repeatability was determined with the coefficient of variation (CoV) and the coefficient of repeatability (CoR) values. Agreement between devices was examined by Bland-Altman analysis and intraclass correlation coefficient (ICC) analyses. RESULTS: The CoV and CoR for IOP values were 18.4% and 5.5 mmHg with the NT530P and 16.6% and 4.3 mmHg with the GAT respectively. On average, NT530P overestimated the IOP compared with GAT by 1.3 mmHg. The 95% limits of agreement between the methods ranged from -5.9 to +3.3 mmHg, and the ICC for average IOP measures between devices was 0.64 (P < 0.001). The CoV and CoR for CCT measurements were 5.5% and 59.8 mum with the NT530P and 5.5% and 60.5 mum with UP. NT530P CCT readings were on average 13 mum thinner than those obtained with UP. The 95% limits of agreement (LoA) ranged from -7.4 to +33.1 mum, and the ICC for average CCT measures between devices was 0.93 (P < 0.001). CONCLUSION: The NT530P demonstrated high IOP and CCT measurement repeatability in the normal population studied. The device tended to slightly overestimate IOP compared to GAT. CCT values provided by NT530P were interchangeable with those of UP.
机译:背景:要检查由Tono-Pachymeter NT530P(日本Nidek)提供的眼内压(IOP)和中央角膜厚度(CCT)测量的可重复性,并评估该设备与Goldmann压平眼压计(GAT)的一致性和超声波测厚法(UP)。方法:62名受试者的62只正常眼被纳入研究。每个设备完成了三个会话的测量,每个会话包含三组测量。重复性由变异系数(CoV)和重复性系数(CoR)值确定。通过Bland-Altman分析和类内相关系数(ICC)分析检查了设备之间的一致性。结果:NT530P的IOP的CoV和CoR分别为18.4%和5.5 mmHg,GAT的IOP值的CoV和CoR分别为16.6%和4.3 mmHg。平均而言,与GAT相比,NT530P高估了IOP 1.3 mmHg。两种方法之间的一致性协议的95%范围在-5.9到+3.3 mmHg之间,并且设备之间的平均IOP措施的ICC为0.64(P <0.001)。使用NT530P进行CCT测量的CoV和CoR为5.5%和59.8毫米,使用UP时为5.5%和60.5毫米。 NT530P CCT读数平均比UP薄13毫米。协议(LoA)的95%范围从-7.4到+33.1毫米不等,并且设备之间的平均CCT措施的ICC为0.93(P <0.001)。结论:NT530P在正常人群中显示出较高的IOP和CCT测量可重复性。与GAT相比,该设备倾向于略微高估IOP。 NT530P提供的CCT值可与UP互换。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号