The influence of different speech processor and hearing aid settings on speech perception outcomes in electric acoustic stimulation patients.

机译:在电声刺激患者中,不同的语音处理器和助听器设置对语音感知结果的影响。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

OBJECTIVE:: Electric acoustic stimulation (EAS) is an increasingly popular means of treating individuals with a steeply sloping mid-to-high frequency hearing loss, who traditionally do not benefit from hearing instruments. These persons often have too much residual hearing to be considered for a cochlear implant. Several studies have demonstrated the ability both to preserve the remaining low-frequency hearing in these individuals, and to provide significant benefit through combining a cochlear implant with a hearing aid to amplify the same ear. These improvements in performance have been especially noted in noise. Often overlooked is that these outcomes may be influenced by the fitting parameters of both the cochlear implant and the hearing aid. DESIGN:: This study assessed four EAS subjects, with a minimum of 1 month's EAS use, on eight different fitting parameters. Sentence testing in different noise levels (+15, +10, and +5 dB SPL) was conducted. Subjects also evaluated each condition using a visual analogue scale. RESULTS:: Results demonstrated that a reduced overlap of cochlear implant and hearing aid amplification produced best results across listening conditions. CONCLUSIONS:: The hearing aid should be fit to a patient-specific modified audiogram at least up to the point where low-frequency hearing is not measurable. The cochlear implant should be fit from a higher frequency point than is standard in patients without residual hearing in the implanted ear, to provide reduced overlap with the amplification provided by the hearing aid. Therefore, a small amount of overlap between the frequency ranges used by the hearing aid and the cochlear implant seems beneficial.
机译:目的:电声刺激(EAS)是一种治疗倾斜度较高的中高频听力损失的人的日益流行的方法,这些人通常无法从听力仪器中受益。这些人的残余听力通常过多,因此无法考虑植入人工耳蜗。多项研究表明,既可以保留这些个体中剩余的低频听力,又可以通过将耳蜗植入物与助听器结合以放大同一只耳朵来提供显着的益处。这些性能的改进在噪音方面尤为明显。通常被忽视的是,这些结果可能会受到人工耳蜗和助听器的拟合参数的影响。设计::这项研究评估了八个不同拟合参数下的四个EAS对象,最少使用了1个月的EAS。在不同的噪声级别(+ 15,+ 10和+5 dB SPL)下进行了句子测试。受试者还使用视觉模拟量表评估了每种情况。结果:结果表明,在聆听条件下,耳蜗植入物和助听器放大物重叠的减少产生了最佳效果。结论:助听器应至少适合不可测量低频听力的点,以适合患者特定的改良听力图。在植入的耳朵中没有残留听力的患者中,耳蜗植入物的安装频率应高于标准频率,以减少与助听器提供的重叠。因此,助听器和人工耳蜗使用的频率范围之间的少量重叠似乎是有益的。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号