In this study, we have developed a new detection method using Si field effect transistor (FET)-type biosensors, which enables the direct monitoring of antigen-antibody binding within very high-ionic-strength solutions such as 1xPBS and human serum. In the new method, as no additional dilution or desalting processes are required, the FET-type biosensors can be more suitable for ultrasensitive and real-time analysis of raw sample solutions. The new detection scheme is based on the observation that the strength of antigen-antibody-specific binding is significantly influenced by the ionic strength of the reaction solutions. For a prostate specific antigen (PSA), in some conditions, the binding reaction between PSA and anti-PSA in a low-ionic strength reaction solution such as 10 μM phosphate buffer is weak (reversible), while that in high-ionic strength reaction solutions such as 1xPBS or human serum is strong.
机译:在这项研究中,我们开发了一种使用Si场效应晶体管(FET)型生物传感器的新检测方法,该方法可以直接监测非常高离子强度的溶液(例如1xPBS和人血清)中的抗原抗体结合。在这种新方法中,由于不需要额外的稀释或脱盐过程,因此FET型生物传感器可能更适合于对原始样品溶液进行超灵敏的实时分析。新的检测方案基于以下观察结果:抗原-抗体特异性结合的强度受到反应溶液离子强度的显着影响。对于前列腺特异性抗原(PSA),在某些情况下,PSA和抗PSA在低离子强度反应溶液(如10μM磷酸盐缓冲液)中的结合反应较弱(可逆),而在高离子强度反应中1xPBS或人血清等溶液很强。