首页> 外文期刊>Experiments in Fluids: Experimental Methods and Their Applications to Fluid Flow >Frequency response correction of hot-film measurements in a turbulent airflow
【24h】

Frequency response correction of hot-film measurements in a turbulent airflow

机译:湍流中热膜测量的频率响应校正

获取原文
获取原文并翻译 | 示例
           

摘要

Turbulence measurements were performed in a fully developed turbulent pipe flow using a wedged-shaped hot-film probe. The dynamic sensitivity of hot-film probes is frequency dependent and cannot be obtained from static calibration. As it is complicated to make a dynamic calibration, a correction method for the turbulent energy spectrum is used instead. The measurements were compared with hot-wire turbulence measurements and show good agreement. [References: 15]
机译:使用楔形热膜探头在完全展开的湍流管道中进行湍流测量。热膜探头的动态灵敏度与频率有关,无法通过静态校准获得。由于进行动态校准比较复杂,因此使用了湍流能谱的校正方法。将该测量结果与热线湍流测量结果进行比较,并显示出良好的一致性。 [参考:15]