Risk of electromagnetic interference induced by dental equipment on cardiac implantable electrical devices

机译:牙科设备在心脏植入式电子设备上诱发电磁干扰的风险

获取原文
获取原文并翻译 | 示例
           

摘要

Patients with cardiac implantable electrical devices should take special precautions when exposed to electromagnetic fields. Proximity to equipment used in clinical dentistry may cause interference. This study evaluated in vitro the risks associated with different types/makes of cardiac devices and types of dental equipment. Six electronic dental tools were tested on three implantable cardioverter defibrillators and three pacemakers made by different manufacturers. Overall, the risk of interference with the pacemakers was 37% lower than with the implantable cardioverter defibrillators. Regarding the types/makes of cardiac devices analysed, that from Boston Scientific had a five-fold greater risk of interference than did that from Biotronik [prevalence ratio (PR) = 5.58]; there was no difference between that from Biotronik and that from Medtronic. Among the dental equipment, the electric pulp tester had the greatest risk of inducing interference and therefore this device was used as the benchmark. The electronic apex locator (PR = 0.29), Periotest M (PR = 0.47), and the ultrasonic dental scaler (PR = 0.59) were less likely to induce interference than the electric pulp tester. The risk was lowest with the electronic apex locator. Pacemakers presented a lower risk of light to moderate interference (PR = 0.63). However, the risk of severe electromagnetic interference was 3.5 times higher with pacemakers than with implantable cardioverter defibrillators (PR = 3.47).
机译:带有心脏植入式电子设备的患者在电磁场中应采取特殊的预防措施。接近临床牙科设备可能会造成干扰。这项研究在体外评估了与不同类型/品牌的心脏设备和牙科设备类型相关的风险。在三个植入式心脏复律除颤器和由不同制造商生产的三个起搏器上测试了六个电子牙科工具。总体而言,起搏器受到干扰的风险比植入式心脏复律除颤器低37%。关于所分析的心脏设备的类型/品牌,波士顿科学公司的干扰发生风险比Biotronik公司高五倍[患病率(PR)= 5.58]; Biotronik和Medtronic的结果没有区别。在牙科设备中,电动牙髓测试仪产生最大干扰的风险,因此该设备被用作基准。电子牙尖定位器(PR = 0.29),Periotest M(PR = 0.47)和超声波洁牙机(PR = 0.59)比电动牙髓测试器更不可能产生干扰。使用电子根尖定位器的风险最低。起搏器对中度干扰的风险较低(PR = 0.63)。但是,起搏器引起的严重电磁干扰的风险是植入式心脏复律除颤器的3.5倍(PR = 3.47)。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号