PURPOSE: To assess the repeatability of endothelial cell density (ECD) measurements by the EM-935 endothelial microscope and their agreement with those of the Konan-Noncon Robo SP 6000 (Noncon Robo) specular microscope. METHODS: We assessed the agreement between automated and semiautomated methods of analysis of the EM-935 and the Noncon Robo by measuring ECD in 40 eyes of 20 healthy subjects and calculating the 95% limits of agreement (LoA) and plotting Bland-Altman graphs. We then evaluated the repeatability of both the methods of the EM-935 by calculating the intraclass correlation coefficients (ICC) and coefficient of variation (CV). RESULTS: The mean ECD of the 40 eyes was 2531+/-244 cells/mm2 for Noncon Robo and 2483+/-159 cells/mm2 for EM-935-automated. The difference between these 2 methods was not statistically significant. The mean ECD for EM-935-semiautomated was 2635+/-190 cells/mm2 and was statistically significantly higher than the 2 other methods (p<0.001). The 95% LoA were -435 to 339 cells/mm2 for the Noncon Robo and EM-935-automated, -230 to 438 cells/mm2 for the Noncon Robo and EM-935-semiautomated, and -347 to 43 cells/mm2 for the EM-935-automated and EM-935-semiautomated. Repeatability was better for EM-935-semiautomated method compared with automated method as expressed by ICC (95% CI) of 0.80 (0.52-0.95) vs 0.50 (0.09-0.84) and coefficient of variation of 2.43% vs 2.85%, respectively. CONCLUSIONS: The EM-935 specular microscope showed better repeatability for the semiautomated mode compared with the automated mode. Although measurement agreement with the Noncon Robo was somewhat better for the semiautomated mode, agreement was only moderate for both methods. This leads us to recommend that these instruments should not be used interchangeably.
机译:目的:评估通过EM-935内皮显微镜进行的内皮细胞密度(ECD)测量的可重复性,以及与Konan-Noncon Robo SP 6000(Noncon Robo)镜面显微镜的一致性。方法:我们通过测量20名健康受试者的40只眼睛的ECD并计算95%的同意限(LoA)并绘制Bland-Altman图,评估了EM-935和Noncon Robo自动化和半自动化分析方法之间的一致性。然后,我们通过计算类内相关系数(ICC)和变异系数(CV)来评估EM-935两种方法的可重复性。结果:40只眼的平均ECD对于Noncon Robo是2531 +/- 244细胞/ mm2,对于EM-935自动化是2483 +/- 159细胞/ mm2。这两种方法之间的差异无统计学意义。半自动化EM-935的平均ECD为2635 +/- 190细胞/ mm2,在统计学上显着高于其他两种方法(p <0.001)。 Noncon Robo和EM-935自动化的95%LoA为-435至339个细胞/ mm2,Noncon Robo和EM-935半自动化的为-230至438细胞/ mm2,以及-347至43个/ mm2 EM-935自动化和EM-935自动化。半自动方法的重复性优于自动方法,ICC(95%CI)分别为0.80(0.52-0.95)和0.50(0.09-0.84),变异系数分别为2.43%和2.85%。结论:与自动模式相比,EM-935镜面显微镜在半自动模式下显示出更好的可重复性。尽管对于半自动模式,与Noncon Robo的测量协议要好一些,但是对于两种方法,协议都只是中等程度。因此,我们建议这些仪器不应互换使用。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号