Test-retest variability in visual field testing using frequency doubling technology.

机译:使用倍频技术进行视野测试的重测变异性。

获取原文
获取原文并翻译 | 示例
       

摘要

PURPOSE. To quantify the magnitude of test-retest variability (TRV) for normal subjects in serial visual fields (VF) using the frequency doubling technology (FDT) instrument. METHODS. Twenty-one healthy adults, aged 23 to 60 years, underwent four serial FDT VF tests, using the full-threshold C-20 program of the Zeiss-Humphrey FDT analyzer, on one randomly chosen eye. The VF tests were spaced 2 to 4 days apart. All subjects performed two preliminary FDT tests in order to minimize any learning effect. Test-retest variability was calculated as the standard deviation of each location's sensitivity value across the four VF tests. RESULTS. Mean TRV (+/-SD) for the entire field was 2.44+/-1.32 dB. Mean TRV (+/-SD) for the superior, inferior, nasal, and temporal hemifields were 2.48+/-1.3, 2.40+/-1.4, 2.40+/-1.3, and 2.48+/-1.3 dB, respectively. Mean TRV (+/-SD) for the foveal location, the 4 central, and the 12 peripheral locations were 2.49+/-1.4, 2.16+/-1.2, and 2.54+/-1.4 dB, respectively. CONCLUSIONS. TRVwas found to be rather uniform across the visual field of the commercially available FDT device, with only a mild, clinically insignificant, effect of both eccentricity and age on TRV. Variability in the FDT VF, for normal subjects, was found to be more uniform than that of both standard and short wavelength automated perimetry. In addition, a strong inverse correlation was found, in normal subjects, between the mean sensitivity and TRV.
机译:目的。要使用倍频技术(FDT)仪器量化连续视野(VF)中正常受试者的重测变异性(TRV)的大小。方法。随机选择一只眼睛,使用Zeiss-Humphrey FDT分析仪的全阈值C-20程序,对23名年龄在23至60岁之间的健康成年人进行了四次连续FDT VF测试。 VF测试间隔2到4天。所有受试者均进行了两项初步的FDT测试,以最大程度地降低学习效果。重新测试的变异性计算为四个VF测试中每个位置的灵敏度值的标准偏差。结果。整个场的平均TRV(+/- SD)为2.44 +/- 1.32 dB。上,下,鼻和颞半场的平均TRV(+/- SD)分别为2.48 +/- 1.3、2.40 +/- 1.4、2.40 +/- 1.3和2.48 +/- 1.3 dB。中心凹部位,4个中央部位和12个周围部位的平均TRV(+/- SD)分别为2.49 +/- 1.4、2.16 +/- 1.2和2.54 +/- 1.4 dB。结论。发现TRV在市售的FDT装置的整个视野​​上是相当均匀的,偏心率和年龄对TRV只有轻微的,临床上微不足道的影响。对于正常受试者,发现FDT VF的变异性比标准和短波自动视野检查法更为均匀。此外,在正常受试者中,平均灵敏度与TRV之间存在强烈的反相关关系。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号