Reproducibility of optic nerve head measurements obtained by optical coherence tomography.

机译:通过光学相干断层扫描获得的视神经头部测量值的可重复性。

获取原文
获取原文并翻译 | 示例
       

摘要

PURPOSE. To evaluate reproducibility of optic nerve head (ONH) morphometry measurements obtained by optical coherence tomography (the fast optical disk protocol of the -Zeiss model 3000 OCT system) in normal and glaucomatous eyes. METHODS. Prospective instrument-evaluation study. ONH measurements for 20 eyes were obtained in eight scan sessions taken during two visits to an ophthalmology clinic (10 normal patients, 10 glaucoma patients, one eye per subject). At every one of the eight sessions for each eye, estimates of eight ONH morphometry variables (see Main outcome measures) were obtained. The first two sessions were performed by two operators, followed by a 30-minute break. The same operators then completed a third and fourth session. This sequence was duplicated on a second visit. Intrasession, intersession, intervisit, and interoperator reproducibility of the eight variables were calculated by the use of a components variance model. Intraclass correlation coefficients (ICC) were used to assess reliability. RESULTS. Vertical integrated rim area, horizontal integrated rim area, disk area, cup area, rim area, cup/disk area ratio, cup/disk horizontal ratio, cup/disk vertical ratio. With the exception of the horizontal integrated rim area and rim area in normal subjects, the factor subject was the most important source of variance for all variables. Reliability values as measured by ICC for normal eyes were above 81%, with the exception of measurements of the horizontal integrated rim area (23.1%), rim area (33.3%), and disk area (64.7%). For glaucomatous eyes all values were above 85%, with the exception of the disk area (68.1%). CONCLUSIONS. ONH measurements obtained using the fast optical disk protocol of the Zeiss 3000 OCT system show good reproducibility, for both normal and glaucomatous eyes.
机译:目的。要评估通过光学相干断层扫描(-Zeiss 3000型OCT系统的快速光盘协议)在正常和青光眼中获得的视神经乳头(ONH)形态测量结果的可重复性。方法。前瞻性仪器评估研究。在两次就诊于眼科诊所期间进行的八次扫描过程中获得了20只眼睛的ONH测量值(10名正常患者,10名青光眼患者,每名受试者一只眼睛)。在每只眼睛的八个疗程中的每个疗程中,都会获得八个ONH形态测量变量的估计值(请参阅主要结果指标)。前两个会话由两名操作员执行,然后休息30分钟。然后,相同的操作员完成了第三和第四节。再次访问时重复此序列。使用组件方差模型计算了八个变量的闭会期间,闭会期间,互访和互操作性的可重复性。类内相关系数(ICC)用于评估可靠性。结果。垂直一体轮辋面积,水平一体轮辋面积,盘面积,杯面面积,轮圈面积,杯/盘面积比,杯/盘水平比,杯/盘垂直比。除了正常受试者的水平综合边缘区域和边缘区域之外,因子主体是所有变量最重要的方差来源。由ICC测量的正常眼睛的可靠性值高于81%,但水平积分眼圈面积(23.1%),眼圈面积(33.3%)和椎间盘面积(64.7%)除外。对于青光眼,除椎间盘面积(68.1%)外,所有值均高于85%。结论。使用Zeiss 3000 OCT系统的快速光盘协议获得的ONH测量结果对于正常和青光眼均具有良好的再现性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号