PURPOSE. To compare the prevalence of visual field loss, the sensitivity distribution, and the size and depth of glaucomatous visual field defects using the standard full threshold (FT) and the Swedish interactive threshold algorithm (SITA) standard (SS) procedures in patients with early or suspected glaucoma. METHODS. Automated perimetry findings were retrospectively evaluated in 53 patients (105 eyes) with early or suspected glaucoma. RESULTS. The number of eyes judged to have glaucomatous visual field loss by SS (48 eyes) was significantly larger than what was found with FT (35 eyes), and 70 eyes were classified as pre-perimetric glaucoma. In these 70 eyes, there were many locations where the sensitivity was significantly higher with SS than with FT (intrasubject difference), and SS had less intersubject variability than FT at most locations. The cumulative decibel scores at the region of glaucomatous defects were larger with SS (206.2+/-103.3 dB) than with FT (162.1+/-87.5 dB) (p=0.02), which indicated that the depth of defects measured by SS was shallower than that by FT. The sizes of defects were significantly larger with SS (11.2+/-5.6) than with FT (9.7+/-5.1) (p<0.05). CONCLUSIONS. Glaucomatous defects were measured as being significantly shallower and larger with SS than with FT. In addition, the prevalence of visual field defect was higher with SS according to some of the criteria for glaucomatous visual field defects. These results might be related to the fact that SS strategy has a lower variability and to the Bayesian statistical properties of the SITA algorithm.
机译:目的。使用标准全阈值(FT)和瑞典互动阈值算法(SITA)标准(SS)程序比较早期或早期患者的视野丧失的发生率,敏感性分布以及青光眼视野缺损的大小和深度怀疑是青光眼。方法。对53例早期或疑似青光眼患者的自动视野检查结果进行回顾性评估。结果。被SS判定为有青光眼视野丧失的眼数(48只眼)明显大于FT(35只眼),并将70只眼归为围手术期青光眼。在这70只眼睛中,有很多位置SS的敏感性明显高于FT(受试者内部差异),并且在大多数位置,SS的受试者间变异性均低于FT。 SS(206.2 +/- 103.3 dB)在青光眼缺陷区域的累积分贝分数大于FT(162.1 +/- 87.5 dB)(p = 0.02),这表明SS测量的缺陷深度为比FT浅。 SS(11.2 +/- 5.6)的缺陷尺寸明显大于FT(9.7 +/- 5.1)(p <0.05)。结论。与FT相比,SS的青光眼缺陷明显变浅和变大。此外,根据青光眼视野缺损的某些标准,SS视野缺损的患病率较高。这些结果可能与SS策略的可变性较低以及SITA算法的贝叶斯统计特性有关。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号