Magnetic resonance imaging compatibility of endoclips.

机译:内窥镜的磁共振成像兼容性。

获取原文
获取原文并翻译 | 示例
           

摘要

BACKGROUND: Endoscopic clipping devices are now available for treatment of GI hemorrhage and microperforations. All commercially available endoclips are labeled as magnetic resonance imaging (MRI) incompatible. No data are available about the actual magnetic field strength at which endoclips are first deflected nor the clinical relevance of the magnetic fields on endoclips used in GI endoscopy. OBJECTIVE: To determine the compatibility of different endoclips with MRI. DESIGN: Prospective observational study. SETTING: Experiment on excised pig tissue in an MRI scanner. INTERVENTIONS: The physical deflection and strength of attraction of endoclips: Resolution Clip, TriClip, QuickClip, and Ethicon Endo-surgery Clip were measured in different positions by using an MRI scanner at a field strength of 1.5 Tesla. Endoclips that demonstrated deflection were attached to a pig stomach and tested for detachment at a 1.5-Tesla MRI field strength. MAIN OUTCOME MEASUREMENTS: Physical deflection and detachment from pig stomach mucosa in an MRI scanner. RESULTS: All endoclips except the one made by Ethicon Endo-surgery demonstrated physical deflection under the tested conditions. The magnetic attraction was strongest for the Resolution Clip (0.7 gauss) compared with the TriClip (1.2 gauss) and the QuickClip (26.8 gauss). Only the Triclip demonstrated detachment from the pig gastric tissue under testing conditions. LIMITATIONS: A pig model and a small number of clips. CONCLUSIONS: The Ethicon Endo-surgery clip is compatible with MRI. All other clips showed deflection in a magnetic field, but only the TriClip demonstrated detachment from gastric tissue, and hence should be considered MRI incompatible.
机译:背景:内窥镜夹持装置现已可用于治疗胃肠道出血和微穿孔。所有市售的内窥镜均标记为不兼容磁共振成像(MRI)。没有关于内窥镜首先偏转的实际磁场强度的数据,也没有关于胃肠镜检查中使用的内窥镜的磁场的临床相关性的数据。目的:确定不同内窥镜与MRI的相容性。设计:前瞻性观察研究。地点:在MRI扫描仪上对切除的猪组织进行实验。干预:内窥镜的物理挠度和吸引力强度:通过使用MRI扫描仪在场强为1.5 Tesla的不同位置测量了Resolution Clip,TriClip,QuickClip和Ethicon Endo-Surgery Clip。将表现出挠度的内窥镜附着在猪胃上,并在1.5 Tesla MRI场强下测试其分离。主要观察指标:MRI扫描仪中猪胃粘膜的物理偏转和脱离。结果:除由Ethicon Endo-Surgery制造的内窥镜外,所有内窥镜在测试条件下均显示出物理变形。与TriClip(1.2高斯)和QuickClip(26.8高斯)相比,Resolution Clip(0.7高斯)的磁吸引力最强。在测试条件下,只有Triclip证明与猪胃组织脱离。局限性:猪模型和少量夹子。结论:Ethicon Endo手术夹与MRI兼容。所有其他夹子在磁场中均显示偏转,但只有TriClip表现出与胃组织分离,因此应将其视为MRI不兼容。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号