首页> 外文期刊>Eye >The Cataract National Dataset electronic multicentre audit of 55,567 operations: when should IOLMaster biometric measurements be rechecked?
【24h】

The Cataract National Dataset electronic multicentre audit of 55,567 operations: when should IOLMaster biometric measurements be rechecked?

机译:Cataract国家数据集电子多中心审核55,567次操作:何时应重新检查IOLMaster生物特征测量?

获取原文
获取原文并翻译 | 示例
           

摘要

PURPOSE: Calculation of intraocular lens (IOL) power for implantation during cataract surgery depends on ocular biometric measurements. The aim of this study was to characterise the normal range of intra- and interindividual variation in axial length (AL) and corneal power (K) when IOLMaster measurements were possible and to derive recommendations as to which outlying measurements merit verification before acceptance. METHODS: The Medisoft electronic patient database contains prospectively collected data conforming to the United Kingdom (UK) Cataract National Dataset on 55,567 cataract operations. From this AL and K information on the 32,556 eyes (14,016 paired) of patients older than 25 years, without corneal pathology, history of intraocular surgery and who had all biometric measurements taken with the Zeiss IOLMaster (Carl Zeiss Meditec) were extracted. R 2.8.1 (R Foundation for Statistical Computing) was used for statistical analysis. RESULTS: Mean age was 76.4 years and 62.0% were female. Mean (95% confidence interval) values for AL, mean K and corneal astigmatism were 23.40 (21.27-26.59) mm, 43.90 (40.94-47.01) D and 1.04 (<2.50) D. Nearly all astigmatism was either with or against the rule. Differences between paired eyes were not statistically significant. 95% individuals had asymmetry of AL and mean K<0.70 mm and 0.92 D, respectively. CONCLUSIONS: On the basis of approximation of the 95% CI above, it is suggested that AL, mean K and keratometric astigmatism measurements outside the ranges 21.30-26.60 mm, 41.00-47.00 D and >2.50 D, respectively, and intraindividual asymmetry of AL >0.70 mm or mean K>0.90 D should be verified before acceptance.
机译:目的:白内障手术期间人工晶状体(IOL)植入度的计算取决于眼部生物特征测量。这项研究的目的是,在可能进行IOLMaster测量时,描述轴长(AL)和角膜屈光力(K)的个体内和个体间变化的正常范围,并得出建议,在接受之前对哪些外围测量值得验证。方法:Medisoft电子患者数据库包含前瞻性收集的数据,这些数据符合关于55,567例白内障手术的英国(UK)白内障国家数据集。从该AL和K信息中,提取出25岁以上,无角膜病理,眼内手术史以及使用Zeiss IOLMaster(卡尔蔡司Meditec)进行了所有生物特征测量的32556眼患者(14016对)。 R 2.8.1(R统计计算基础)用于统计分析。结果:平均年龄为76.4岁,女性为62.0%。 AL,平均K和角膜散光的平均值(95%置信区间)值为23.40(21.27-26.59)mm,43.90(40.94-47.01)D和1.04(<2.50)D。几乎所有散光均符合或不符合规则。配对眼睛之间的差异无统计学意义。 95%的人的AL不对称,平均K <0.70 mm和0.92D。结论:基于上述95%CI的近似值,建议AL,平均K和角膜散光测量分别在21.30-26.60 mm,41.00-47.00 D和> 2.50 D范围之外,以及AL的个体内不对称性> 0.70 mm或平均K> 0.90 D,应在验收之前进行验证。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号