Reliability of tonosafe disposable tonometer prisms: clinical implications from the Veterans Affairs Boston Healthcare System Quality Assurance Study.

机译:Tonosafe一次性眼压计棱镜的可靠性:退伍军人事务波士顿医疗系统质量保证研究的临床意义。

获取原文
获取原文并翻译 | 示例
           

摘要

PURPOSE: Given the Veterans Affairs Boston Healthcare System's recent introduction of single-use Tonosafe disposable tonometer prisms as an alternative to Goldmann applanation tonometers (GATs), this study had two aims: to conduct a large-scale quality assurance trial to assess the reliability of intraocular pressure (IOP) measurements of the Tonosafe disposable tonometer compared with GAT, particularly at extremes of pressure; to evaluate the suitability of Tonosafe disposable tonometer prisms as an acceptable substitute for GATs and for clinic-wide implementation in an academic tertiary referral setting. METHODS: Ophthalmology resident physicians measured the IOPs of patients in general and specialty eye clinics with the Tonosafe disposable tonometer and GAT. Tonosafe test-retest reliability data were also collected. A retrospective review of patient charts and data analysis were performed to determine the reliability of measurements. RESULTS: The IOPs of 652 eyes (326 patients) were measured with both GAT and Tonosafe, with a range of 3-34 mm Hg. Linear regression analysis showed R=0.93, slope=0.91, both of which supported the proposed hypothesis, and the y-intercept=-1.05 was significantly different from the hypothesized value. The Tonosafe test-retest repeatability (40 eyes of 40 patients), r=0.977, was very high, which was further supported by linear regression slope=0.993, y-intercept=0.118, and a Tonosafe repeatability coefficient of 2.06, similar to GAT repeatability. CONCLUSIONS: The IOP measurements by Tonosafe disposable prisms correlated closely with Goldmann measurements, with similar repeated measurement variability to GAT. This suggests that the Tonosafe is an acceptable substitute for GAT to measure IOP in ophthalmology clinic settings.
机译:目的:鉴于退伍军人事务波士顿医疗系统最近推出了一次性使用的Tonosafe一次性眼压计棱镜,以替代高曼压平眼压计(GAT),该研究有两个目的:进行一项大规模的质量保证试验,以评估其可靠性。与GAT相比,Tonosafe一次性眼压计的眼内压(IOP)测量,尤其是在极端压力下;评估Tonosafe一次性眼压计棱镜作为GAT可接受的替代品以及在学术第三级转诊环境中在诊所范围内实施的适用性。方法:眼科住院医师使用Tonosafe一次性眼压计和GAT测量普通和专科眼科诊所患者的IOP。还收集了Tonosafe重测可靠性数据。回顾性检查患者图表和数据分析,以确定测量的可靠性。结果:GAT和Tonosafe均测量了652眼(326例患者)的IOP,范围为3-34 mm Hg。线性回归分析显示R = 0.93,斜率= 0.91,两者均支持所提出的假设,并且y截距= -1.05与假设值存在显着差异。 Tonosafe重测重复性(40眼40眼),r = 0.977,很高,线性回归斜率= 0.993,y轴截距= 0.118,Tonosafe可重复性系数2.06,与GAT相似,进一步证明了这一点。重复性。结论:Tonosafe一次性棱镜的IOP测量值与Goldmann测量值密切相关,并且重复测量的变化与GAT相似。这表明在眼科临床环境中,Tonosafe可以替代GAT来测量IOP。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号