Varying effects of corneal thickness on intraocular pressure measurements with different tonometers.

机译:不同眼压计对角膜厚度对眼内压测量的影响。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

AIMS: To compare the intraocular pressure (IOP) measurements using the Goldmann applanation tonometer (GAT), noncontact tonometer (NCT), and ocular blood flow tonometer (OBFT), and to evaluate the effects of varying central corneal thickness (CCT) on the readings. METHODS: Ultrasound pachymetry and tonometry were performed on 170 eyes. Using the corrected GAT values that took CCT into account as the standard, we calculated the NCT and OBFT measurement errors related to the CCT variable by a linear regression model. Group comparisons were performed with the chi2 test and one-way ANOVA test. The correlation between the various tonometer measurements was analysed by Pearson's correlation method. RESULTS: Both the NCT (r=0.872, P<0.001) and OBFT measurements (r=0.861, P<0.001) were highly correlated with the GAT measurements. IOP measurements using the three tonometers were all correlated with CCT (all P<0.001), with the NCT measurements showing the greatest regression coefficient (beta=0.063, r=0.650) andthe GAT measurements the least (beta=0.037, r=0.496). A linear regression model indicated that a 10 microm change in CCT resulted in a NCT measurement deviation of 0.47-0.98 mmHg and an OBFT measurement deviation of 0.29-0.81 mmHg. CONCLUSION: Pressure readings with the GAT, NCT, and OBFT are all affected by CCT, with the NCT being the one most affected and the GAT the least. Our findings suggest CCT an essential variable to consider in interpreting IOP readings, especially for the NCT measurements.
机译:目的:比较使用戈德曼压平眼压计(GAT),非接触式眼压计(NCT)和眼血流量眼压计(OBFT)进行的眼内压(IOP)测量,并评估改变中央角膜厚度(CCT)对阅读。方法:对170只眼进行超声测厚和眼压测定。使用将CCT考虑在内的校正后的GAT值作为标准,我们通过线性回归模型计算了与CCT变量相关的NCT和OBFT测量误差。使用chi2检验和单向ANOVA检验进行组比较。通过皮尔逊相关法分析了各种眼压计测量值之间的相关性。结果:NCT(r = 0.872,P <0.001)和OBFT测量(r = 0.861,P <0.001)均与GAT测量高度相关。使用三个眼压计的IOP测量值均与CCT相关(所有P <0.001),NCT测量值显示最大回归系数(β= 0.063,r = 0.650),而GAT测量值最小(β= 0.037,r = 0.496)。 。线性回归模型表明,CCT变化10微米会导致NCT测量偏差为0.47-0.98 mmHg,OBFT测量偏差为0.29-0.81 mmHg。结论:GCT,NCT和OBFT的压力读数均受CCT影响,其中NCT受影响最大,而GAT最少。我们的发现表明,CCT是解释IOP读数时要考虑的重要变量,尤其是对于NCT测量而言。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号