The composition of protective aprons worn by X-ray personnel to shield against secondary radiation is changing. Lead is being replaced by either lead-free or composite (lead with other high atomic numbered elements) materials. These newer aprons are categorised by manufacturers in terms of lead equivalent values, but it is unclear how these stated values compare with actual lead equivalent values. In this work, the actual lead equivalence of 41 protective aprons from four manufacturers, all specified as having 0.25 mm lead equivalence, were investigated with transmission experiments at 70 and 100 kVp. All aprons were in current use. The aprons were screened for defects, and age, weight and design was recorded along with details of associated quality assurance (QA). Out of the 41 protective aprons examined for actual lead equivalence, 73% were outside tolerance levels, with actual levels in some aprons demonstrating less than half of the nominal values. The lack of compatibility between actual and nominal lead equivalent values was demonstrated by aprons from three of the four manufacturers investigated. The area of the defects found on screening of the protective aprons were within recommendations. The results highlight the need for acceptancy and ongoing checks of protective aprons to ensure that radiation exposure of imaging personnel is kept to a minimum.
机译:X射线人员为防止二次辐射而佩戴的防护围裙的组成正在改变。铅被无铅或复合材料(含其他具有高原子序数的元素的铅)材料代替。制造商根据铅当量值对这些较新的围裙进行了分类,但尚不清楚这些规定值与实际铅当量值如何比较。在这项工作中,通过在70 kVp和100 kVp下的传输实验研究了来自四个制造商的41个防护围裙的实际铅当量,均被指定为0.25 mm铅当量。目前所有围裙都在使用中。筛选围裙的缺陷,并记录年龄,体重和设计以及相关质量保证(QA)的详细信息。在检查的41个防护围裙的实际铅当量中,有73%超出了公差范围,某些围裙的实际围护水平还不到标称值的一半。被调查的四个制造商中有三个制造商的围裙证明了实际和名义上的铅当量值之间缺乏兼容性。在筛选保护性围裙时发现的缺陷区域在建议范围内。结果突出表明需要对防护围裙进行接受和持续检查,以确保将成像人员的辐射暴露保持在最低水平。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号