Orthodontic springs and auxiliary appliances: assessment of magnetic field interactions associated with 1.5 T and 3 T magnetic resonance systems.

机译:正畸弹簧和辅助器具:评估与1.5 T和3 T磁共振系统相关的磁场相互作用。

获取原文
获取原文并翻译 | 示例
           

摘要

The objective of this paper is to evaluate magnetic field interactions at 1.5 and 3 T for 20 orthodontic devices used for fixed orthodontic therapy. Twenty springs and auxiliary parts made from varying ferromagnetic alloys were tested for magnetic field interactions in the static magnetic field at 1.5 and 3 T. Magnetic translational force F(z) (in millinewtons) was evaluated by determining the deflection angle beta [American Society for Testing and Materials (ASTM standard test method)]. Magnetic-field-induced rotational force F(rot) was qualitatively determined using a five-point scale. beta was found to be >45 degrees in 13(15) devices at 1.5(3) T and translational force F(z) exceeded gravitational force F(g) on the particular object [F(z) 10.17-261.4 mN (10.72-566.4 mN) at 1.5(3) T]. F(z) was found to be up to 24.1(47.5)-fold higher than F(g) at 1.5(3) T. Corresponding to this, F(rot) on the objects was shown to be high at both field strengths (> or = +3). Three objects (at 1.5 T) and one object (at3 T) showed deflection angles <45 degrees , but F(rot) was found to be > or +3 at both field strengths. For the remaining objects, beta was below 45 degrees and torque measurements ranged from 0 to +2. Of 20 objects investigated for magnetic field interactions at 1.5(3) T, 13(15) were unsafe in magnetic resonance (MR), based on the ASTM criteria of F(z). The implications of these results for orthodontic patients undergoing MRI are discussed.
机译:本文的目的是评估20种用于固定正畸治疗的正畸设备在1.5和3 T下的磁场相互作用。测试了二十种由不同铁磁合金制成的弹簧和辅助零件在1.5和3 T静磁场中的磁场相互作用。通过确定偏转角β来评估磁平移力F(z)(以毫牛顿为单位)。测试和材料(ASTM标准测试方法)]。使用五点标度定性确定磁场感应的旋转力F(rot)。发现在1.5(3)T的13(15)装置中,β大于45度,并且特定物体上的平移力F(z)超过重力F(g)[F(z)10.17-261.4 mN(10.72-在1.5(3)T时为566.4 mN)。发现F(z)在1.5(3)T时比F(g)高24.1(47.5)倍。与此相对应,在两个场强下,物体上的F(rot)都很高( >或= +3)。三个物体(在1.5 T时)和一个物体(在3 T时)显示的偏转角<45度,但在两种场强下,F(rot)均大于或等于+3。对于其余物体,β低于45度,扭矩测量范围为0至+2。根据F(z)的ASTM标准,在1.5(3)T下进行磁场相互作用研究的20个对象中,有13(15)个在磁共振(MR)中不安全。讨论了这些结果对正畸MRI患者的意义。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号