The objective of this study was to compare the radiation exposure delivered by helical CT and pulmonary angiography (PA) for the detection of pulmonary embolism (PE), with an anthropomorphic phantom. A preliminary survey defined a representative standard procedure for helical CT and PA ( n=148) by choosing the exposure settings most frequently used. Then, radiation doses were measured with thermoluminescent dosimeters TLD 100 (Lif) introduced into the depth of an anthropomorphic phantom. Average doses were approximately five times smaller with helical CT than with PA (6.4+/-1.5 and 28+/-7.6 mGy, respectively). The most important doses were abreast the pulmonary apex for CT, and abreast the pulmonary arteries for PA. Compared with PA, helical CT dose distribution was relatively uniform (10-13 mGy). Finally, concerning abdomen and pelvis, doses were more important for PA than for CT scan (0.06-2.86 and 0.2-11.5 mGy, respectively). For the diagnostics of PE, radiation exposure is five times smaller with helical CT than with pulmonary angiography.
机译:这项研究的目的是将螺旋CT和肺血管造影(PA)所提供的辐射暴露与拟人化体模进行比较,以检测肺栓塞(PE)。初步调查通过选择最常用的曝光设置为螺旋CT和PA(n = 148)定义了代表性的标准程序。然后,用引入拟人体模深度的热辐射剂量计TLD 100(Lif)测量辐射剂量。螺旋CT的平均剂量大约是PA的五倍(分别为6.4 +/- 1.5和28 +/- 7.6 mGy)。最重要的剂量是CT并发于肺尖,PA并发于肺动脉。与PA相比,螺旋CT的剂量分布相对均匀(10-13 mGy)。最后,关于腹部和骨盆,PA的剂量比CT扫描更重要(分别为0.06-2.86和0.2-11.5 mG​​y)。对于PE的诊断,螺旋CT的放射线暴露比肺血管造影小五倍。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号